| Device | T:SLIM G4 INSULIN PUMP WITH DEXCOM G4 PLATINUM CGM |
| Generic Name | Sensor, glucose, invasive |
| Applicant | Tandem Diabetes Care, Inc. 11045 Roselle St. San Diego, CA 92121 |
| PMA Number | P140015 |
| Supplement Number | S007 |
| Date Received | 12/30/2015 |
| Decision Date | 01/28/2016 |
| Product Codes |
MDS OYC |
| Advisory Committee |
Clinical Chemistry |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
| Recalls | CDRH Recalls |
Approval Order Statement CHANGES TO THE PACKAGING VERIFICATION PROCESSES TO REPLACE 100% VISUAL INSPECTION WITH 100% PACKAGE CONTENTS VERIFICATION USING VALIDATED CHECKWEIGHT SCALE, AND CHANGES TO THE LABELING VERIFICATION PROCESS TO REPLACE 100% VISUAL VERIFICATION WITH ONCE-DAILY VISUAL INSPECTION OF A RETAINED LABEL FOR THE T:SLIM G4 INSULIN PUMP WITH DEXCOM G4 PLATINUM CGM SYSTEM. |