| Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
| |
| Device | ANIMAS VIBE SYSTEM |
| Generic Name | Pump, infusion, insulin, to be used with invasive glucose sensor |
| Applicant | Animas Corp. 965 Chesterbrook Blvd. Wayne, PA 19087 |
| PMA Number | P130007 |
| Supplement Number | S013 |
| Date Received | 04/12/2016 |
| Decision Date | 05/12/2016 |
| Withdrawal Date | 06/11/2020 |
| Product Codes |
MDS OYC |
| Advisory Committee |
Clinical Chemistry |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
| Recalls | CDRH Recalls |
Approval Order Statement Change to implement a new fixture, new set-up configuration, and a control software modification for the purpose of increasing throughput for one of the incoming inspection processes used for lot acceptance of the insulin cartridges of the Animas Vibe Insulin Pump. The Animas Vibe Insulin Pump is a component of the Animas Vibe System. |
|
|