| Device | MINIMED 530G INSULIN PUMP |
| Generic Name | Automated insulin dosing , threshold suspend |
| Applicant | Medtronic, Inc. 18000 Devonshire St. Northridge, CA 91325 |
| PMA Number | P120010 |
| Supplement Number | S095 |
| Date Received | 09/19/2016 |
| Decision Date | 10/19/2016 |
| Product Code |
OZO |
| Advisory Committee |
Clinical Chemistry |
| Clinical Trials | NCT01112696
|
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
| Recalls | CDRH Recalls |
Approval Order Statement Add a new testing station to conduct a manual Displacement Accuracy Test (DAT) in the manufacturing process of the Paradigm Real-Time Insulin Pumps, Paradigm Real-Time Revel Pumps, and MiniMed 530G Insulin Pumps. These pumps are components of the Paradigm Real-Time System, Paradigm Real-Time Revel System, and MiniMed 530G System, respectively. The testing station will be added to the manufacturing facility, located at Juncos, Puerto Rico. |