| Device | PARADIGM REAL-TIME REVEL SYSTEM |
| Generic Name | Sensor, glucose, invasive |
| Applicant | Medtronic Minimed 18000 Devonshire St. Northridge, CA 91325 |
| PMA Number | P150019 |
| Supplement Number | S024 |
| Date Received | 11/09/2016 |
| Decision Date | 01/25/2017 |
| Product Codes |
MDS OYC |
| Advisory Committee |
Clinical Chemistry |
| Clinical Trials | NCT01464346
|
| Supplement Type | Real-Time Process |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
| Recalls | CDRH Recalls |
Approval Order Statement Approval for adding the requirement for a chemical exposure durability test to the pump product specifications and for a manufacturing process change of adding an annealing process during the assembly of the pump cases. |