| Device | Medtronic MiniMed 630G Insulin Pump |
| Generic Name | Automated insulin dosing , threshold suspend |
| Applicant | Medtronic Minimed 18000 Devonshire St. Northridge, CA 91325 |
| PMA Number | P150001 |
| Supplement Number | S023 |
| Date Received | 09/14/2017 |
| Decision Date | 10/11/2017 |
| Product Code |
OZO |
| Advisory Committee |
Clinical Chemistry |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
| Recalls | CDRH Recalls |
Approval Order Statement New manufacturing process to allow printed circuit board assemblies to be harvested and re-used for construction of refurbished and loaner 630G and 670G insulin pumps. The 630G and 670G insulin pumps are components of the MiniMed 630G and MiniMed 670G systems, respectively. |