| |
| Device | MiniMed 630G System with SmartGuard |
| Generic Name | Automated insulin dosing , threshold suspend |
| Applicant | Medtronic Minimed 18000 Devonshire St. Northridge, CA 91325 |
| PMA Number | P150001 |
| Supplement Number | S026 |
| Date Received | 10/10/2017 |
| Decision Date | 11/07/2017 |
| Product Code |
OZO |
| Advisory Committee |
Clinical Chemistry |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
| Recalls | CDRH Recalls |
Approval Order Statement New electroplating machine to increase manufacturing capacity for the Guardian Sensor (3) continuous glucose monitoring sensor. The Guardian Sensor (3) is a component of the MiniMed 630G and MiniMed 670G systems. |