| Device | MiniMed 630G Pump |
| Generic Name | Automated insulin dosing , threshold suspend |
| Applicant | Medtronic Minimed 18000 Devonshire St. Northridge, CA 91325 |
| PMA Number | P150001 |
| Supplement Number | S033 |
| Date Received | 01/23/2018 |
| Decision Date | 02/22/2018 |
| Product Code |
OZO |
| Advisory Committee |
Clinical Chemistry |
| Supplement Type | Special (Immediate Track) |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
| Recalls | CDRH Recalls |
Approval Order Statement Approval for the addition of 100% visual inspection to the manufacturing protocol for P-cap O-rings (Part Number D5022027-001) and addition of a second de-flashing procedure to remove excess flash in out-of-spec (OOS) O-rings having flash >0.003. |