• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Devices@FDA

  • Print
  • Share
  • E-mail
-
 
Help | Download Files | More about PMA

New Search Back to Search Results
Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceMiniMed 670G Pump
Generic NameAutomated insulin dosing device system, single hormonal control
ApplicantMedtronic Minimed, Inc.
18000 Devonshire St.
Northridge, CA 91325
PMA NumberP160017
Supplement NumberS030
Date Received01/23/2018
Decision Date02/22/2018
Product Code OZP 
Advisory Committee Clinical Chemistry
Clinical TrialsNCT02130284
Supplement TypeSpecial (Immediate Track)
Supplement Reason Process Change - Manufacturer/Sterilizer/Packager/Supplier
Expedited Review Granted? No
Combination ProductNo
Predetermined Change Control Plan AuthorizedNo
RecallsCDRH Recalls
Approval Order Statement  
Approval for addition of 100% visual inspection to the manufacturing protocol for P-cap O-rings (Part Number D5022027-001) and addition of a second de-flashing procedure to remove excess flash in out-of-spec (OOS) O-rings having flash >0.003.
Post-Approval StudyShow Report Schedule and Study Progress
  We welcome your comments and feedback about Devices@FDA.
-
-