| Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
| |
| Device | MiniMed 630G System |
| Generic Name | Automated insulin dosing , threshold suspend |
| Applicant | Medtronic Minimed 18000 Devonshire St. Northridge, CA 91325 |
| PMA Number | P150001 |
| Supplement Number | S070 |
| Date Received | 06/13/2019 |
| Decision Date | 07/12/2019 |
| Product Code |
OZO |
| Advisory Committee |
Clinical Chemistry |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
| Recalls | CDRH Recalls |
Approval Order Statement Addition of a new final functional tester for the motor assembly component. The motor assembly is a component of the MiniMed 630G and MiniMed 670G systems. |
|
|