| Device | MiniMed 670G System |
| Generic Name | Automated insulin dosing device system, single hormonal control |
| Applicant | Medtronic Minimed, Inc. 18000 Devonshire St. Northridge, CA 91325 |
| PMA Number | P160017 |
| Supplement Number | S108 |
| Date Received | 01/26/2023 |
| Decision Date | 04/18/2023 |
| Product Code |
OZP |
| Advisory Committee |
Clinical Chemistry |
| Clinical Trials | NCT02130284
|
| Supplement Type | Real-Time Process |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
| Recalls | CDRH Recalls |
Approval Order Statement Approval for a minor design change to the Next Generation Pumps (NGP) to remove remote bolus functionality, remove remote settings modification, and disable over the air programming. The proposed changes address vulnerabilities in connection with Field Action (FA1272) and will be implemented in the MiniMed 670G and 630G Insulin Pumps. |
| Post-Approval Study | Show Report Schedule and Study Progress |