Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 046 Date of Entry 12/23/2016 
FR Recognition Number 14-498
Standard
ASTM F2097-16 Standard Guide for Design and Evaluation of Primary Flexible Packaging for Medical Products
Scope/Abstract
1.1 This guide provides directions for the design and evaluation of primary flexible packages for medical products. The package materials must be selected appropriately for manufacturing process, end use, and the product being packaged.

1.2 This guide provides a compendium of test methods, practices, and procedures. Specific individual test methods must be selected based on the pertinent characteristics of the specific product to be packaged and the purpose for testing, research and development, or compliance. Not all test methods will be applicable.

1.3 This guide does not address acceptability criteria, which need to be determined jointly by the package producer and the medical products manufacturer.

Extent of Recognition
Complete standard
Relevant FDA Guidance and/or Supportive Publications*
ANSI/AAMI/ISO 11607-1:2006(R2010) Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems [Including: Amendment 1 (2014)].

ANSI/AAMI/ISO 11607-2:2006(R2010) Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly process [Including: Amendment 1 (2014)].

ANSI/AAMI/ISO TIR16775:2014 Packaging for terminally sterilized medical devices - Guidance on the application of ISO 11607-1 and ISO 11607-2.
FDA Technical Contact
 Patrick Weixel
  FDA/OC/CDRH/OPEQ/OIDRH/DRH/
  301-796-5537
  Patrick.Weixel@fda.hhs.gov
Standards Development Organization
ASTM ASTM International http://www.astm.org/
FDA Specialty Task Group (STG)
Sterility
*These are provided as examples and others may be applicable.
-
-