U.S. flag An official website of the United States government
  1. Home
  2. Drug Databases
  3. Drug Safety-related Labeling Changes

Drug Safety-related Labeling Changes (SrLC)

Get Email Alerts | Guide

RUBY-FILL (NDA-202153)

(RUBIDIUM CHLORIDE RB-82)

Safety-related Labeling Changes Approved by FDA Center for Drug Evaluation and Research (CDER)

Download Data

Expand all

04/26/2019 (SUPPL-12)

Approved Drug Label (PDF)

Boxed Warning

Revised; Full text below:

High Level Radiation Exposure with Use of Incorrect Eluent

Patients are exposed to high radiation levels when the generator is eluted with the incorrect eluent due to high Sr 82 and Sr 85 breakthrough levels.

  • Use only additive-free 0.9% Sodium Chloride Injection USP to elute the generator.
  • Immediately stop the patient infusion and discontinue the use of the affected RUBY-FILL (Rb 82) generator if the incorrect solution is used to elute the generator.
  • Evaluate the patient’s radiation absorbed dose and monitor for the effects of radiation to critical organs such as bone marrow.

 

Excess Radiation Exposure with Failure to Follow Quality Control Testing Procedure

Excess radiation exposure occurs when the levels of Sr 82 or Sr 85 in the rubidium Rb 82 chloride injection exceed specified limits.

  • The system automatically generates a record and saves the data for each generator eluate volume, including flushing and test volumes. Total cumulative eluate volumes are also recorded and saved for the life of the generator.

  • Strictly adhere to the generator quality control testing procedure, to minimize the risk of excess radiation exposure, including daily testing and additional testing at Alert Limits.

  • Stop use of a generator at any of the following Expiration Limits. Expiry Limits are:

    • 30 L for the generator's cumulative eluate volume, or

    • Expiration date of the generator (60 days post-manufacturing)

    • An eluate Sr 82 level of 0.01 µCi /mCi (kBq/MBq) Rb 82, or

    • An eluate Sr 85 level of 0.1 µCi /mCi (kBq/MBq) Rb 82.

4 Contraindications

Addition of the following:

RUBY-FILL is contraindicated for use if a solution other than additive-free 0.9% Sodium Chloride Injection USP has been used to elute the generator at any time. Immediately stop the patient infusion and permanently discontinue the use of the affected RUBY-FILL generator whenever the incorrect eluent is used.

5 Warnings and Precautions

Revised Subsection titles underlined below:

5.1 High Level Radiation Exposure with Use of Incorrect Eluent

Newly added information as below:

Use only additive free 0.9% Sodium Chloride Injection USP to elute the generator. Apply the provided saline tag to the additive free 0.9% Sodium Chloride Injection USP container before use. Additives present in other solutions (particularly calcium ions) expose patients to high levels of radiation by causing the release of large amounts of Sr 82 and Sr 85 into the eluate regardless of the generator’s age or prior use.

Immediately stop the patient infusion and discontinue use of the affected RUBY-FILL generator if the incorrect eluent is used and evaluate the patient’s radiation absorbed dose and monitor for the effects of radiation to critical organs such as bone marrow. When solutions containing calcium ions are used to elute the generator, high levels of radioactivity are present in any subsequent eluate, even with the use of additive free 0.9% Sodium Chloride Injection USP.

5.2 Excess Radiation Exposure with Failure to Follow Eluate Testing Protocol

Additions and/or revisions underlined:

Excess radiation exposure occurs when the Sr 82 and Sr 85 levels in rubidium Rb 82 chloride injections exceed the specified generator eluate limits.

Strictly adhere to the eluate testing protocol to minimize radiation exposure to the patient. Stop using the rubidium generator when the expiration limits are reached.

6 Adverse Reactions

Additions and/or revisions underlined:
Radiation Exposure

High level radiation exposure to the bone marrow has occurred in some patients due to Sr 82 and Sr 85 breakthrough in the eluate when an incorrect solution was used to elute the rubidium Rb 82 generator Excess radiation exposure has occurred …