Model Number TJF-Q190V
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code Drug Resistant Bacterial Infection (4553)
Date of Event 07/21/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that there were four (4) potential patients directly impacted by one suspected medical device.The patients underwent procedures involving the same individual suspected medical device between (b)(6) 2024.The patient referenced in this report was admitted to the hospital on (b)(6) 2024 for cholecystitis and septic shock.The patient's blood cultures were positive for pan sensitive e.Coli on admission.The patient underwent an endoscopic retrograde cholangiopancreatography (ercp) procedure with suspected medical device to remove stones on (b)(6) 2024.The patient's gallbladder drain was cultured on (b)(6) 2024 and was positive for carbapenem resistant e.Coli (cp-cre).The cp-cre was not treated since the patient was improving on current regimen.The patient was discharged on home health on (b)(6) 2024.On (b)(6) 2024, ercp procedure involving suspected medical device was identified as possible link between multiple cp-cre infections.On (b)(6) 2024, suspected medical device was examined by borescope during facility investigation.Borescope revealed a small tear within the inner sheath of the device.It is unknown when the tear appeared on the scope, what caused it, and whether it was present for ercp procedures involving affected patients.The customer reiterated that the device had an inner sheath tear, which the customer was not aware of, and they used the scope on several patients.The customer further stated that the device might have caused a cross contamination between patients.Reportedly, the device has been in quarantine in a secure location.The customer confirmed that the device passed the leak test prior to procedures being done.This report is related to the following linked patient identifiers: patient 1 (b)(6), patient 2 (b)(6), patient 3 (b)(6), patient 4 (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III DUODENOVIDEOSCOPE
Common Device Name
DUODENOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
MDR Report Key20345557
Report Number2429304-2024-0000418
Device Sequence Number20043756
Product Code FDT
UDI-Device Identifier04953170405563
UDI-Public04953170405563
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 10/01/2024
Report Date (Section F) 09/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTJF-Q190V
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date09/05/2024
Event Location Hospital
Date Report to Manufacturer09/05/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/01/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXTRACTOR PRO RX 7-6F 15-18MM 200CM; HYDRATOME RX 4.4F 20MM W/ 260CM GW; RX ACCESSORY RAPID REFILL CONT LNJ SYSTEM; SPYSCOPE DS II DIGITAL ACCESS & DELIVERYCATH 4661; SYPGLASS RETRIEVAL BASKET 286CM X 15MM
Outcome Attributed to Adverse Event Other;
Patient Age80 YR
Patient SexFemale
Patient Weight79 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-