The customer complained of discrepant inr results for 1 patient from caoguchek xs meter serial number (b)(4) compared to the laboratory result.At 06:31 am, the result from the meter was 5.0 inr.The result from the laboratory was 2.9 inr.The specific time of the laboratory collection was not provided.There was no allegation of an adverse event.The patient was "healthy".The qc was acceptable.The patient was not anemic, had no heparin, no antiphospholipid antibodies, and no lupus.The patient's therapeutic range was 2 - 3 inr.The suspect product was requested to be returned for investigation.Retention test strips (lot 286319) were tested in comparison to master lot #28632180.For this purpose, two human blood samples from marcumar donors and two internal reference meters were used.No error messages occurred.Retention material complies with specification.Extensive measurements have shown higher deviations for coaguchek values > 4.5 inr compared to a laboratory method.Therefore, the customer allegation has been substantiated.A product problem has been found for coaguchek values > 4.5 inr, which was the case in this complaint.This can be traced back to the calibration of the complained test strips to the who standard rtf/16.For measurements > 4.5 inr.The customer is advised to contact the physician and perform a measurement with a laboratory method.Roche diagnostics has issued a recall for this issue.Please refer to medwatch field recall - correction/removal report no.
|
The customers's meter and strips were returned for investigation.Testing results: vial#1 qc 1: 2.4 inr, qc 2: 2.5 inr, qc 3: 2.4 inr.Vial #2 qc 1: 2.5 inr, qc 2: 2.4 inr.Retention strips: qc 3: 2.5 inr.The obtained qc values were in the allowed range of the used combination master lot strip lot - qc lot.1.9 - 2.9 inr.All measurements were without error messages.The maximum difference between measurements was 4.0 %.No information was provided in the complaint case that would point to a cause for the result discrepancy.Extensive measurements have shown higher deviations for coaguchek values > 4.5 inr compared to a laboratory method.Therefore, the customer allegation has been substantiated.A product problem has been found for coaguchek values > 4.5 inr, which was the case in this complaint.This can be traced back to the calibration of the complained test strips to the who standard rtf/16.For measurements > 4.5 inr.The customer is advised to contact the physician and perform a measurement with a laboratory method.Roche diagnostics has issued a recall for this issue.Correction/removal report no.
|