Sechrist field technician found the chamber emergency decompression was out of specification, at 125 seconds.Field technician checked the exhaust screen and determined the cause to be a clogged lint on the inside of the exhaust screen.Replaced the exhaust screen and retested the chamber emergency decompression.The chamber decompressed within specification at 80 seconds.At this time there is no evidence that a manufacturing non-conformity contributed to the reported complaint, and the instructions for use were reviewed and determined to provide adequate instructions and warnings for the safe and effective use of the device.Therefore, no corrective or preventative actions are necessary.All complaints are trended and reviewed by management on a monthly basis.As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.Chambers are multi-use, serviceable items.As such, it is anticipated that the units may occasionally require repair, and as part of standard care the chambers should be inspected prior to use.If damage or functional issues are noted during these routine chamber inspections, the unit should not be put into use with a patient and should be reported to sechrist for servicing.Manufacturer reference file #(b)(4).
|