• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE PS RP INSRT SZ 5 8MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE PS RP INSRT SZ 5 8MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Model Number 1516-50-508
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Impaired Healing (2378)
Event Date 01/06/2022
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the patient had a knee hardware removal and placement of antibiotic spacer due to poor healing of original surgical site and outer skin quality.Doi: (b)(6) 2021.Dor: (b)(6) 2022, affected side: right knee.
 
Manufacturer Narrative
Product complaint # (b)(4) investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS RP INSRT SZ 5 8MM
Type of Device
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13282159
MDR Text Key283958955
Report Number1818910-2022-01100
Device Sequence Number1
Product Code NJL
UDI-Device Identifier10603295052456
UDI-Public10603295052456
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P830055
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1516-50-508
Device Catalogue Number151650508
Device Lot Number9800474
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ATTUNE MEDIAL DOME PAT 35MM.; ATTUNE PS FEM RT SZ 5 NAR CEM.; ATTUNE PS RP INSRT SZ 5 8MM.; ATTUNE RP TIB BASE SZ 4 CEM.
Patient Outcome(s) Required Intervention;
Patient Age63 YR
Patient SexFemale
-
-