• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC.1818910 ATTUNE RP TIBIAL IMPACTOR; KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC.1818910 ATTUNE RP TIBIAL IMPACTOR; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 254401004
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 03/23/2015
Event Type  malfunction  
Event Description
Attune impactor broke.
 
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Conclusion and justification status: the complaint states attune impactor broke.The investigation confirmed that the impactor had broken as reported.Expert opinion indicates that the failures are associated with environmental stress cracking (esc).The attune rp tibial impactor has been annealed to reduce residual stresses.Lab tests indicate significant reduction in residual stresses of annealed samples in comparison to un-annealed samples however a complete reduction in stresses is not achievable.This device is from an annealed batch.A field safety notice was issued in nov 2014 stating to reduce the possibility of leaving fragments in patients, the company suggests adhering to the instructions for use (ifu), which include inspecting the impactor¿s to ensure that no instruments or pieces of instruments are left in the surgical site prior to closure.It should be noted that capa-(b)(4) has been initiated and can be referenced for further details.The complaint shall be closed with a justified conclusion; it will be entered into the complaint database and monitored through trend analysis.
 
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP TIBIAL IMPACTOR
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
steve dowell
700 orthopaedic drive
warsaw, IN 46581
5743714918
MDR Report Key4638017
MDR Text Key5576022
Report Number1818910-2015-17392
Device Sequence Number1
Product Code LXH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254401004
Device Lot NumberAU3650772
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/16/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/05/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age69 YR
Patient Weight95
-
-