1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC ATTUNE FB CR ARTIC SURF SZ6; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 254500526
Device Problems Break (1069); Component Missing (2306)
Patient Problem No Information (3190)
Event Type  malfunction  
Event Description
Attune articulating surface is cracked, balseal post is sheared off, and is missing a balseal spring.
 
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Additional narrative: the investigation remains closed, as the added information does not change the outcome of the investigation.
 
Manufacturer Narrative
Examination of the returned device confirms the reported event of trial damage and breakage.A complaints review was completed in february 2014 and in june 2014 for all attune trial instruments, following 25,000 and 50,000 surgeries, inclusive of the attune femoral trials, articulation surface trials, and the patella trials.Findings of the complaints review prompted dfmea updates where applicable.The root cause is attributed to user technique and/or misuse.The damage suggests the trial was pryed or otherwise inappropriately removed during trialing.Based on the root cause of suspected misuse, corrective action is not needed.Monitor subsequent reports via sep-419.Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB CR ARTIC SURF SZ6
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
1219655 DEPUY-RAYNHAM, A DIV. OF DEPUY ORTHOPAEDIC
325 paramount drive
raynham MA 02767
Manufacturer Contact
steve dowell
700 orthopaedic drive
warsaw, IN 46581
5743714918
MDR Report Key4645851
MDR Text Key5600118
Report Number1818910-2015-17599
Device Sequence Number1
Product Code JWH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK140881
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor,distributor,health prof
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500526
Device Lot NumberMVMBRT050
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received08/12/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/13/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-