Catalog Number 254401004
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 05/14/2015
Event Type  malfunction  
Event Description
During implanting the attune tibial component with the attune tibia rp impactor, the impactor broke down to two pieces.
 
Manufacturer Narrative
Additional narrative: this complaint is still under investigation.Depuy will notify the fda of the results of the investigation once it has been completed.
 
Manufacturer Narrative
This complaint is under investigation.Depuy will notify the fda of the results of the investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Additional narrative: the complaint states during implanting the attune tibial component with the attune tibia rp impactor, the impactor broke down to two pieces.The investigation confirmed that the impactor had broken as reported.Expert opinion indicates that the failures are associated with environmental stress cracking (esc).The attune rp tibial impactor has been annealed to reduce residual stresses.Lab tests indicate significant reduction in residual stresses of annealed samples in comparison to un-annealed samples however a complete reduction in stresses is not achievable.This device is from an annealed batch.A field safety notice was issued in nov 2014 stating to reduce the possibility of leaving fragments in patients, the company suggests adhering to the instructions for use (ifu), which include inspecting the impactor¿s to ensure that no instruments or pieces of instruments are left in the surgical site prior to closure.It should be noted that capa-(b)(4) has been initiated and can be referenced for further details.The complaint shall be closed with a justified conclusion; it will be entered into the complaint database and monitored through trend analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP TIBIAL IMPACTOR
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
steve dowell
700 orthopaedic drive
warsaw, IN 46581
5743714918
MDR Report Key4790196
MDR Text Key5853532
Report Number1818910-2015-21378
Device Sequence Number1
Product Code LXH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 05/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/22/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254401004
Device Lot NumberAU3664789
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/16/2015
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-