Model Number MMT-754CMB |
Device Problem
Self-Activation or Keying (1557)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/12/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.
The device will be returned for analysis and further information will follow once the analysis has been completed.
No conclusion can be drawn at this time.
The insulin pump involved in this event is the paradigm real-time veo insulin infusion pump, which is not marketed in the united states.
However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|
|
Event Description
|
It was reported that the numbers on the insulin pump scroll, without any input from the customer.
Customer's blood glucose level at the time was unknown.
Troubleshooting occurred.
Advised to discontinue use of the insulin pump, and revert to a back-up plan.
The customer was advised that the device would be replaced and agreed to return the product for analysis.
|
|
Manufacturer Narrative
|
The insulin pump was received with intermittent button response due to moisture damage on the keypad trace.
No frozen display noted and the numbers do not ramp up on its own or scroll bar moving with no input.
The insulin pump has minor scratched lcd window, cracked case near the display window corners, battery tube threads cracked and cracked reservoir tube lip.
|
|