MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-754WWB
Device Problem Device Displays Incorrect Message (2591)
Patient Problem Hyperglycemia (1905)
Event Date 07/13/2015
Event Type  malfunction  
Manufacturer Narrative
Currently, it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the paradigm real-time veo insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
 
Event Description
It was reported the insulin pump didn't deliver a basal and the insulin pump didn't alarm no delivery.Customer was hospitalized twice for high blood glucose of 25 mmol/l.Customer current blood glucose is 5.5 mmol/l.Test was performed and the device did alarm no delivery.The device is being returned for further analysis.
 
Manufacturer Narrative
A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.However, the pump had a cracked case at the display window corner, cracked battery tube threads and a cracked reservoir tube lip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
christopher tupper
18000 devonshire street
northridge, CA 91325-1219
8185764313
MDR Report Key4946274
MDR Text Key23097056
Report Number3004209178-2015-78183
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Followup
Report Date 07/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-754WWB
Device Catalogue NumberMMT-754WWB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received07/14/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-