• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. ATTUNE KNEE SYSTEM(MB); KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. ATTUNE KNEE SYSTEM(MB); KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 254401006
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 11/23/2015
Event Type  malfunction  
Manufacturer Narrative
This complaint is under investigation.Depuy will notify the fda of the results of the investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
Attune impactor is broken.
 
Manufacturer Narrative
This complaint remains under investigation.Depuy will notify the fda of the results of the investigation once it has been completed.
 
Manufacturer Narrative
Conclusion and justification status: the complaint states during the inspection of attune returned loan kits, the orthokit technician noted that: 2 attune impactors had cracks; 2 attune impactors were broken.The investigation confirmed that 2 of the impactors had broken and 2 had cracked with fragments missing.Expert opinion indicates that the failures are associated with environmental stress cracking (esc).The attune impactors have been annealed to reduce residual stresses.Lab tests indicate significant reduction in residual stresses of annealed samples in comparison to un-annealed samples however a complete reduction in stresses is not achievable.A field safety notice was issued in nov 2014 stating to reduce the possibility of leaving fragments in patients, the company suggests adhering to the instructions for use (ifu), which include inspecting the impactor¿s to ensure that no instruments or pieces of instruments are left in the surgical site prior to closure.The complaint shall be closed with a justified conclusion; it will be entered into the complaint database and monitored through trend analysis.Depuy still considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE KNEE SYSTEM(MB)
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key5266973
MDR Text Key32671091
Report Number1818910-2015-36642
Device Sequence Number1
Product Code HWA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 12/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254401006
Device Lot NumberAU3870946
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received01/14/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/24/2014
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-