Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problems
Hyperglycemia (1905); Hypoglycemia (1912)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging the basal history did not match the programmed basal rates.It was reported the patient's blood glucose ranged from below 70 mg/dl and above 40 mg/dl to above 250 mg/dl and below 500 mg/dl with no signs or symptoms of hypoglycemia and hyperglycemia.The reported health event does not qualify as a serious injury.This complaint is being reported because the reported issue was not resolved with troubleshooting.
|
|
Manufacturer Narrative
|
Product analysis: the device was returned and evaluated by product analysis on 05/17/2016 with the following findings: the complaint could not be duplicated or confirmed with investigation.Review of the pump¿s black box revealed: a loss of prime on (b)(6) 2016 at 12:38 associated with a cartridge change; deliveries resumed at 16:35; a loss of prime following a rewind step on (b)(6) 2016 at 21:55.The pump was manually suspended on the same date at 22:17 follow by a rebooting event at 22:47.The pump powered on (b)(6) 2016 at 06:22; deliveries did not resume.The total daily dose history was appropriate for the user programmed delivery settings.The pump successfully completed a prime sequence, bolus deliveries, and 24-hour exercise test without issue or alarm.The pump¿s history properly recorded all deliveries performed during investigation.
|
|
Search Alerts/Recalls
|