• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE SPACER BLOCK; ATTUNE INSTRUMENTS : SPACER BLOCKS/TRIAL SHIMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE SPACER BLOCK; ATTUNE INSTRUMENTS : SPACER BLOCKS/TRIAL SHIMS Back to Search Results
Catalog Number 254401014
Device Problem Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/06/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the attune spacer block and attune spacer base will not stay together and need replacement.
 
Manufacturer Narrative
Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint (b)(4).Investigation summary : conclusion and justification status: the complaint states it was reported that the attune spacer block and attune spacer base will not stay together and need replacement.The devices were reviewed and no product problem was identified.The products (spacer block and spacer base) fit together.The complaint shall be closed with a unjustified conclusion; entered into the complaints database and monitored through trend analysis.If further information is received the complaint shall be reopened and investigated further.Post market surveillance is per sep 419.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE SPACER BLOCK
Type of Device
ATTUNE INSTRUMENTS : SPACER BLOCKS/TRIAL SHIMS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork IN
EI 
MDR Report Key7163850
MDR Text Key96388808
Report Number1818910-2018-50191
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295395379
UDI-Public10603295395379
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup
Report Date 12/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254401014
Device Lot NumberBFA0VPN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2018
Date Manufacturer Received10/11/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-