• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND 9616671 ATTUNE CONV SHIM SZ3 16MM; ATTUNE INSTRUMENTS : TIBIAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND 9616671 ATTUNE CONV SHIM SZ3 16MM; ATTUNE INSTRUMENTS : TIBIAL TRIALS Back to Search Results
Catalog Number 254500919
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation summary: examination of the returned device confirms the reported event of trial damage.The observed cracking is initiating around the metal insert of the trial.A pra (b)(4) (preliminary risk assessment) meeting was held to discuss a product escalation out of (b)(6) concerning the attune shims.The team has determined that there is no additional patient risk associated with this issue.Although cracks are occurring, this failure mode does not lead to a patient harm of infection.The rate of infection for attune is within the state rate in the risk management report and is statistically similar to the rate of infection for the total knee class.Although a definitive root cause is not known, a combination user technique, improper technique or misuse, and the use of contraindicated chemicals (non-compliant to cleaning guidelines), may lead to weakening of the material, as seen in the returned device.Based on the performed investigation and pra (b)(4) determination of no additional patient risk, corrective action is not indicated.Continue to monitor via sep-419.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that instruments were broken.Didn't impact any case.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned device confirms the reported event of trial damage.The observed cracking is initiating around the metal insert of the trial.A pra (b)(4) (preliminary risk assessment) meeting was held to discuss a product escalation out of australia concerning the attune shims.The team has determined that there is no additional patient risk associated with this issue.Although cracks are occurring, this failure mode does not lead to a patient harm of infection.The rate of infection for attune is within the state rate in the risk management report and is statistically similar to the rate of infection for the total knee class.Although a definitive root cause is not known, a combination user technique, improper technique or misuse, and the use of contraindicated chemicals (non-compliant to cleaning guidelines), may lead to weakening of the material, as seen in the returned device.Based on the performed investigation and pra 103121982 determination of no additional patient risk, corrective action is not indicated.Continue to monitor via sep-419.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CONV SHIM SZ3 16MM
Type of Device
ATTUNE INSTRUMENTS : TIBIAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46582-0988
5743725905
MDR Report Key7673340
MDR Text Key113529388
Report Number1818910-2018-64098
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295134961
UDI-Public10603295134961
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500919
Device Lot NumberBFA0QBS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-