Model Number 15
Device Problem Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 06/19/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio replaced the therapy cable which resolved the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.
 
Event Description
Physio-control was performing preventative maintenance on the customer's device and observed that the therapy cable was intermittently connecting to the device.It was a very loose connected and the device would lose the ecg signal if the cable was moved.As a result, defibrillation therapy may not be available, if it was needed.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 15 DEFIBRILLATOR/MONITOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
megan marseglia
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key7679897
MDR Text Key113747091
Report Number0003015876-2018-01085
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142430
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number15
Device Catalogue Number99577-001255
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/04/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-