Catalog Number 254401004
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/03/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation summary: the investigation finds the impactor is cracked.Expert opinion indicates that the failures are associated with environmental stress cracking (esc).The attune rp tibial impactor has been annealed to reduce residual stresses.Lab tests indicate significant reduction in residual stresses of annealed samples in comparison to un-annealed samples however a complete reduction in stresses is not achievable.This device is from an annealed batch.A field safety notice was issued in nov 2014 stating to reduce the possibility of leaving fragments in patients, the company suggests adhering to the instructions for use (ifu), which include inspecting the impactor¿s to ensure that no instruments or pieces of instruments are left in the surgical site prior to closure.It should be noted that capa-(b)(4) has been initiated and can be referenced for further details.The complaint shall be closed with a justified conclusion; it will be entered into the complaint database and monitored through trend analysis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that instrument broke while impacting tibial tray.All pieces were recovered and no delay in the case.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP TIBIAL IMPACTOR
Type of Device
ATTUNE INSTRUMENTS : IMPACTORS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic dr.
warsaw, IN 46582
5743725905
MDR Report Key7725738
MDR Text Key115207259
Report Number1818910-2018-65319
Device Sequence Number1
Product Code HWA
UDI-Device Identifier10603295130208
UDI-Public10603295130208
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254401004
Device Lot NumberAU5352178
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-