• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND 9616671 ATTUNE IMPACTION HANDLE; ATTUNE INSTRUMENTS : HANDLES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND 9616671 ATTUNE IMPACTION HANDLE; ATTUNE INSTRUMENTS : HANDLES Back to Search Results
Catalog Number 254401017
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/06/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint : (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the impaction handle spring broke when surgeon was impacting femur.No surgical delay.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> evaluation of the returned device finds the locking feature is fractured.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE IMPACTION HANDLE
Type of Device
ATTUNE INSTRUMENTS : HANDLES
Manufacturer (Section D)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer Contact
kara ditty-bovard
1210 ward avenue
west chester, PA 19380
6103142063
MDR Report Key8451732
MDR Text Key139837423
Report Number1818910-2019-88470
Device Sequence Number1
Product Code HWA
UDI-Device Identifier10603295423324
UDI-Public10603295423324
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254401017
Device Lot NumberNW147389
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/18/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/14/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age71 YR
-
-