• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE RP TIBIAL IMPACTOR; ATTUNE INSTRUMENTS : IMPACTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE RP TIBIAL IMPACTOR; ATTUNE INSTRUMENTS : IMPACTORS Back to Search Results
Catalog Number 254401004
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/21/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that while implanting the attune tibial component it appeared the tibial impactor was broken.Needs to be replaced.No time lost and all pieces accounted for and will be sent back.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision depuy synthesof 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : examination of the returned device confirmed the reported event.Root cause and corrective action are documented in the capa system.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthese is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.Product complaint # : (b)(4).Investigation summary: the device associated with this report was not returned for evaluation, thus the reported event could not be confirmed.The investigation could not verify or identify any product contribution to the reported event with the information provided.Depuy synthese considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP TIBIAL IMPACTOR
Type of Device
ATTUNE INSTRUMENTS : IMPACTORS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key8785084
MDR Text Key150848517
Report Number1818910-2019-98168
Device Sequence Number1
Product Code HWA
UDI-Device Identifier10603295130208
UDI-Public10603295130208
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup
Report Date 06/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/12/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254401004
Device Lot NumberAU6325809
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2019
Date Manufacturer Received11/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age62 YR
-
-