Device rfid chip for dental appliance
Regulation Description Implantable radiofrequency transponder system for patient identification and health information.
Definition The device is intended to enable access to secure patient identification and device information when used with complete dentures, partial dentures, and other removable oral appliances.
Physical State Solid state RFID transponder embedded in a resin
Technical Method Uses RFID transponder and NFC to transmit information to a mobile device.
Target Area Embedded in removable oral appliances
Regulation Medical Specialty General Hospital
Review Panel General Hospital
Product CodePYQ
Premarket Review Drug Delivery and General Hospital Devices, and Human Factors (DHT3C)
Drug Delivery and General Hospital Devices, and Human Factors (DHT3C)
Submission Type 510(K) Exempt
Regulation Number 880.6300
Device Class 2
Total Product Life Cycle (TPLC) TPLC Product Code Report
GMP Exempt? No
Summary Malfunction
Reporting
Eligible
    Note: Class II devices the Food and Drug Administration (FDA) has also published a list of class II (special controls) devices subject to certain limitations, that are exempt from premarket notification requirements under the Food and Drug Administration Modernization Act of 1997 (FDAMA) and the 21st Century Cures Act of 2016 (Cures Act). FDA believes that these exemptions will relieve manufacturers from the need to submit premarket notification submissions for these devices and will enable FDA to redirect the resources that would be spent on reviewing such submissions to more significant public health issues. FDA is taking this action in order to meet requirements of FDAMA and the Cures Act.
Implanted Device? No
Life-Sustain/Support Device? No
Third Party Review Not Third Party Eligible
-
-