Device |
one-way air-leak valve |
Definition |
To restrict air flow to the target section of the lung. To control prolonged air leaks of the lung, or significant air leaks that are likely to become prolonged air leaks, following lobectomy, segmentectomy, or lung volume reduction surgery (lvrs). An air leak present on post-operative day 7 is considered prolonged unless present only during forced exhalation or cough. An air leak present on day 5 should be considered for treatment if it is: 1) continuous, 2) present during normal inhalation phase of inspiration, or 3) present upon normal expiration and accompanied by subcutaneous emphysema or respiratory compromise. |
Physical State |
This device is a small nitinol frame that has a polyurethane membrane over the Nitinol frame that acts as a plug to the targeted section of the lung. |
Technical Method |
The device is a nitonal frame with a polyurethane membrane over the Nitinol frame. The valve does not allow air into the targeted segment of the lung, but does allow air and mucus out of that lung section. |
Target Area |
The device is intended to occlude a section of lung. |
Review Panel |
Anesthesiology |
Product Code | OAZ |
Premarket Review |
Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices
(OHT1)
Division of Anesthesia, Respiratory, and Sleep Devices
(DHT1C)
|
Submission Type |
HDE - Humanitarian Device Exemption
|
Device Class |
HDE
|
Total Product Life Cycle (TPLC) |
TPLC Product Code Report
|
GMP Exempt? |
No
|
Summary Malfunction Reporting |
Eligible |
Implanted Device? |
Yes
|
Life-Sustain/Support Device? |
No
|
Third Party Review |
Not Third Party Eligible |
|
|