The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

Long Term Registry


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Completed
Application Number /
Requirement Number
P030050 / PAS001
Date Original Protocol Accepted 08/17/2009
Date Current Protocol Accepted 08/17/2009
Study Name Long Term Registry
Device Name SCULPTRA AND SCULPTRA AESTHETIC
General Study Protocol Parameters
Study Design Other Study Design
Data Source New Data Collection
Comparison Group No Control
Analysis Type Descriptive
Study Population Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives This study is an five year, open-label, multi-center study to evaluate the long-term safety of Sculptra in HIV-infected patients with facial lipoatrophy by Fitzpatrick skin type and by gender, which will be assessed by the incidence and severity of concomitant illnesses (adverse events). Each subject will be enrolled into one of the four groups at baseline: (a) female with Fitzpatrick skin types I-III; (b) female with Fitzpatrick skin types IV-VI; (c) male with Fitzpatrick skin types I-III; and (d) male with Fitzpatrick skin types IV-VI. Patients will be treated with Sculptra (4-6 treatment sessions to reach optimal correction) at 4 to 6 week intervals.
Study Population This device is indicated for use in immune-competent subjects as a single regimen for correction of shallow to deep nasolabial fold contour deficiencies and other facial wrinkles. The study population includes patients with HIV-associated lipoatrophy.
Sample Size 300 patients, 20 sites
Key Study Endpoints Effectiveness endpoints are the clinical presentation of lipoatrophy (James scale) and physician and subject satisfaction on the device based on questionnaire and photographic comparison before and after the previous visit. Safety endpoints include (1) Incidence and severity of concomitant illnesses and the usage and indications of concomitant medications at baseline, yearly follow-up visits, and unscheduled visits that are not mandated by protocol; (2) Incidence and severity of AEs, including hypertrophic scars or keloids at the yearly follow-up visits and unscheduled visits that are not mandated by protocol; (3) Information regarding any additional Sculptra administered at the yearly and unscheduled visits that are not mandated by protocol. In addition, subject-reported QOL/body image will be collected validated QOL and body image questionnaires.
Follow-up Visits and Length of Follow-up Subjects will have a follow-up visit at one year post baseline (approx 6 months post last treatment), and then yearly office visits for a total of 5 years
Interim or Final Data Summary
Interim Results Preliminary data has shown Sculptra to have a reasonable effectiveness and safety profile at approximately two year after initial treatment.
Actual Number of Patients Enrolled 296
Actual Number of Sites Enrolled 18
Patient Follow-up Rate 81.4%
Final Safety Findings 87.2% of subjects reported treatment emergent adverse events (TEAEs) during the 5�]year study. The majority of TEAEs were mild or moderate in severity (19.0% and 42.1% respectively). The most frequently reported SOC categories were: Infections and Infestations (60.0%), General Disorders and Administrative Site Conditions (43.1%), and Gastrointestinal Disorders (34.1%). One subject reported a TEAE of hypertophic scar or keloid.

Seventy six subjects (26.2%) reported serious TEAEs and 17 subjects (5.9%) died during study. None of the SAEs or deaths was related to either the injection procedure or study medication.
Final Effect Findings There was statistically significant improvement in clinical presentation of facial lipoatrophy at all follow�up time�points (all p<0.001). Significant reductions in mean James scale scores at 1 Year (�-1.4) was either maintained or slightly reduced for the remainder of follow�]up (�-1.4, �-1.1, �-1.1, and �-1.1 at 2, 3, 4, and 5 Years, respectively). Overall, the proportion of subjects with Grade 1 facial lipoatrophy (almost normal) increased from baseline (11.5%) to Year 1 (81.3%) and was maintained or slightly reduced for the remainder of follow�up (79.2%, 61.1%, 58.3%, and 60.1% for the 2, 3, 4, and 5 Year time�points, respectively). Mean change in James Scale scores from baseline by gender and Fitzpatrick skin type were consistent with the overall results.
Recommendations for Labeling Changes Update labeling to include 5 year data from the post approval study


Long Term Registry Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
Semi Annual PAS Report 05/17/2007 05/17/2007 On Time
Semi Annual Report 12/19/2007 12/19/2007 On Time
3 year report 11/10/2008 11/10/2008 On Time
4 year report 11/10/2009 11/06/2009 On Time
60 month report 11/10/2010 11/03/2010 On Time
72 month report 11/10/2011 11/03/2011 On Time
Final Report 03/12/2012 03/12/2012 On Time