The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

Urolume PAS


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Completed
Application Number /
Requirement Number
P920023 / PAS001
Date Original Protocol Accepted 05/06/1996
Date Current Protocol Accepted  
Study Name Urolume PAS
Device Name UROLUME ENDOPROSTHESIS
General Study Protocol Parameters
Study Design Prospective Cohort Study
Data Source New Data Collection
Comparison Group No Control
Analysis Type Descriptive
Study Population Transit.Adolescent B(as adults): 18 yrs < 22 yrs, Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives The objectives were to assess the incidence, duration and severity of positive urine cultures, incontinence and post-void dribbling, to assess changes in sexual dysfunction and global function, and to assess the number of retreatments.
Study Population Study population is as per device indication. Urinary stent indicated for use in men with recurrent bulbal urethral stricture, prostatic obstruction secondary to benign prostatic hyperplasia (BPH), or detrusor external sphincter dyssynergia over age 60 or under age 60 who are poor surgical candidates and who have prostate glands at least 2.0cm in length.
Sample Size 110
Key Study Endpoints Incidence, frequency and severity of the following: urinary tract infection (measured by culture), incontinence, and post-void dribbling. Changes in sexual function, global assessment and number of retreatments
Follow-up Visits and Length of Follow-up At 6-weeks, 6-months, 1year and 2year post-stent placement
Interim or Final Data Summary
Actual Number of Patients Enrolled 72 enrolled but 61 patients treated
Actual Number of Sites Enrolled 10
Patient Follow-up Rate 78.7
Final Safety Findings Negative urine culture at 2 years: 76.2% (95% CI 71.7% - 80.7%). Stress incontinence: 18.5% (10/54) at baseline and 72.2% (39/54) at final evaluation (either at 1- or 2-year visit). Patient bothered by stress incontinence: 31% (13/42) not bother at 2 years vs. 81.4% (48/59) at baseline (p-value<0.001). Urge incontinence: 42.6% (23/54) at baseline and 59.3% (32/54) at final evaluation. Patient bothered by urge incontinence: 48.8% (20/41) not bother at 2 years vs. 55.9% (33/59) at baseline (p-value>0.05). Overflow incontinence: 16.7% (9/54) at baseline and 46.3% (25/54) at final evaluation. Patient bothered by overflow incontinence: 64.3% (27/42) not bother at 2 years vs. 83.1% (49/59) at baseline (p-value<0.05).
Post void dribbling: 63% (34/54) at baseline and 72.2% (39/54) at final evaluation. Patient bothered by post void dribbling: 35.7% (15/42) not bother at 2 years vs. 33.9% (20/59) at baseline (p-value<0.05).
Patients’ responses indicated no negative effects because of stent placement on sexual function. Global assessment (If you were to spend the rest of your life with your urinary condition just the way it is now, how would you feel about that?): Roughly 55.7% of patients reported new satisfaction with their urinary status.
The probability of patients remaining free from re-treatment for recurrent strictures or reobstruction within the stented area in the study was greater than 90% through the 1-year evaluation and approx. 87% at the 2-year follow-up.
Study Strengths & Weaknesses Low precision of estimate since a 72-patient study has a 81% precision and a CI that is 124% wider compared to the planned 110-patients study.
Recommendations for Labeling Changes As per PAS closing letter: Labeling should be revised to include the long term follow-up, safety information and adverse events, including information regarding the proportion of explants observed during the post-approval study, which his twice that identified in the literature


Urolume PAS Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
Follow-up final report on PAS patients 05/30/2007 05/30/2007 On Time