The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

Prolieve PAS


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Completed
Application Number /
Requirement Number
P030006 / PAS001
Date Original Protocol Accepted 02/19/2004
Date Current Protocol Accepted 03/11/2005
Study Name Prolieve PAS
Device Name PROLIEVE THERMODILATATION SYSTEM
General Study Protocol Parameters
Study Design Prospective Cohort Study
Data Source New Data Collection
Comparison Group No Control
Analysis Type Descriptive
Study Population Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives This is an open label non-randomized prospective study . The primary objective of the study is to provide descriptive data on the long-term safety and effectiveness of Prolieve for Benign Prostate Hyperplasia . A secondary objective is to assess safety and effectiveness of re-treatment and to assess the proportion of patients electing re-treatment with Prolieve compared to alternative therapy (other localized therapy, surgery or drug therapy).
Study Population Study population is as per device indication. Adult men men with prostate volume of 20 to 80 grams and in whom drug therapy (Proscar) is typically indicated.
Sample Size The protocol proposes to enroll 250 new subjects in up to 30 study sites in the United States. The protocol states this will ensure at least 100 subjects will be followed for 5 years. In addition to newly enrolled subjects, the sponsor planned to encourage Pivotal Study participants and Continued Access Study participants to transfer to the postmarket study for the completion of 5 years of follow-up
Key Study Endpoints Efficacy is evaluated using the following: Time to re-treatment, American Urological Association (AUA) Symptom Index, (Classifies the severity of subject symptoms related to urinary difficulties based on seven questions that form a total score. The score can range from 0-35). Pain/Discomfort, Sexual Function, Impact of Lower Urinary Tract Symptoms (LUTS) on quality of life (QOL), (Six questions related to urinary problems and how they interfere with daily living. The score ranges from 5-24), Quality of Life Questionnaire, (Six questions related to the subjects feelings about his urinary condition, perception of urinary difficulties, sexual function, activities of daily living, general well-being and social activities. The seven-point scale ranges from delighted to terrible and scores can range from 6-42), Brief Symptom Inventory (BSI), (Seven questions related to the interference of the subject s urinary problems with common activities, scores range from 7-35), BPH Impact Index (BII), (Four questions related to concern about urinary problems and the amount of physical discomfort experienced. Scores can range from 4-17. Safety is measured by: Use of pain medication pre-treatment and during treatment, Pain associated with treatment, Local and systemic symptoms during Prolieve treatment
Catheterizations associated with treatment, PSA levels, Adverse events, Anticipated adverse events (associated with treatment or BPH), Hospitalizations , and MDR reportable events.
Follow-up Visits and Length of Follow-up Study participants are to be evaluated at two weeks, three months and annually thereafter for a period of 5 years after treatment with Prolieve.
Interim or Final Data Summary
Actual Number of Patients Enrolled 225 subjects were enrolled.
Actual Number of Sites Enrolled 24 siteswere enrolled.
Patient Follow-up Rate 102/219 (46.6%) at five year follow-up.
Final Safety Findings There were nine serious device/treatment-relatedadverse events
that required hospitalization. The most common device/treatment-related AEswere micturition urgency (13.3 events per 100 person-years), dysuria (10.5), haematuria (7.8), urinary retention (7.2), bladder
spasm (6.1), pollakiuria (5.7), urethral hemorrhage (5.2), and urine flow decrease (3.1).
Final Effect Findings The retreatment rate was 51.1% (95% CI: 43.2, 59.0) at five years. Subjects showed a general improvement in symptoms; AUA (American Urological Association Symptom Index) total scores, peak flow rate, and quality of life scores all improved from baseline to five years. A smaller percentage of patients reported pain or discomfort in urethra at five years (10/98, 10.2%) than at baseline (55/220, 25.0%).
Study Strengths & Weaknesses Strengths: Prospective design with five-year follow-up.
Weaknesses: The follow-up rate was moderately low (46.6%). However, the study met the assumptions of the protocol to obtain five year follow-up data for at least 100 subjects. Additional sensitivity analyses did not show significant differences between subjects who completed follow-up and those who did not.
Recommendations for Labeling Changes Labeling change is recommended to reflect the results of the five year follow-up post-approval study. The labeling change should include a new section on the label showing a summary of the post-approval study design, results, and study strengths and limitations.


Prolieve PAS Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
2 year report 02/18/2006 03/20/2006 Overdue/Received
3 year report 02/18/2007 07/02/2007 Overdue/Received
4 year report 02/18/2008 02/19/2008 On Time
5 year report 02/17/2009 02/18/2009 Overdue/Received
6 year report 02/19/2010 02/19/2010 On Time
7 year report 02/22/2011 02/22/2011 On Time
72 month report 02/18/2013 12/21/2012 On Time
7 year report (83 month) 04/18/2014 04/21/2014 Overdue/Received
8 year report 05/18/2015 05/14/2015 On Time
9 year report 06/18/2016 06/21/2016 Overdue/Received
10 year report/final report 06/02/2017 06/15/2017 Overdue/Received