| General |
| Study Status |
Completed |
Application Number / Requirement Number |
P980003 / PAS001 |
| Date Original Protocol Accepted |
02/02/1999
|
| Date Current Protocol Accepted |
 
|
| Study Name |
PAS
|
| Device Name |
CHILLI COOLED RF ABLATION SYSTEM
|
| General Study Protocol Parameters |
| Study Design |
Prospective Cohort Study
|
| Data Source |
New Data Collection
|
| Comparison Group |
Objective Performance Criterion
|
| Analysis Type |
Analytical
|
| Study Population |
Adult: At least 22 yrs
|
| Detailed Study Protocol Parameters |
| Study Objectives |
This was a one-armed, observational, prospective cohort study, the purpose of which was to compare device safety and effectiveness with objective performance criteria.
|
| Study Population |
The study population was patients for whom the Chilli device is indicated, who had failed or were intolerant to drug therapy.
|
| Sample Size |
200 patients, 19 sites
|
| Key Study Endpoints |
Effectiveness Endpoints: ventricular tachycardia recurrence rate, defined as the percent of subjects with one or more episodes of ventricular tachycardia recurrence within six months following the index procedure. Safety Endpoints: Assessed as the occurrence rate of a procedure-related major adverse event within seven days of the index procedure.
|
| Follow-up Visits and Length of Follow-up |
Endpoints were assessed at 6-months post-implant
|
| Interim or Final Data Summary |
| Actual Number of Patients Enrolled |
182 patients
|
| Actual Number of Sites Enrolled |
19 sites
|
| Patient Follow-up Rate |
88%
|
| Final Safety Findings |
The primary effectiveness goal was not met. The success criteria was a ventricular tachycardia recurrence rate of 54.2%, whereas the upper 95% confidence limit was 63.8%
|
| Study Strengths & Weaknesses |
Strength: Propensity analysis was conducted to control for confounding variables; Weakness: It is unclear whether to OPCs, on which the success criteria were based, were appropriate.
|
| Recommendations for Labeling Changes |
Labeling changes were made to reflect the failure to meet part of the success criteria.
|