The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

Registry


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Completed
Application Number /
Requirement Number
P010015 / PAS001
Date Original Protocol Accepted 08/28/2001
Date Current Protocol Accepted 08/28/2001
Study Name Registry
Device Name MEDTRONIC INSYNC(TM) BIVENTRICAL PACING SYSTEM
General Study Protocol Parameters
Study Design Prospective Cohort Study
Data Source Sponsor Registry
Comparison Group No Control
Analysis Type Descriptive
Study Population Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives This is a non-randomized multi-center registry. The purpose of this study is to characterize the postmarket performance of the InSync Model 8040 (InSync), InSync III Model 8042 (InSync III), Attain LV leads (Models 2187, 2188, OTW 4193, OTW 4194), InSync ICD Model 7272, InSync Marquis Model 7277, InSync II Marquis Model 7289, InSync Maximo Model 7303, and InSync Sentry Model 7297.
Study Population Study Populaton: Patients from the completed Multicenter InSync Randomized Clinical Evaluation (MIRACLE), U.S. InSync III trials, and patients implanted after market release approval of the InSync, InSync III CRT, and CRT-D systems may be enrolled into the InSync Registry.
Sample Size Up to 2000 patients may be enrolled to obatin 1000 patients with 3 years of follow-up, 150 sites
Key Study Endpoints The primary endpoints include 1) mortality and cause specific mortality, 2) Left ventricular sensing: biventricular pacing impedance, biventricular R wave amplitude (sensing), and LV pacing voltage threshold, and 3) lead-related complications.
Follow-up Visits and Length of Follow-up Clinical data is collected at baseline and then every 6 months for 36 months or until explant or death.
Interim or Final Data Summary
Actual Number of Patients Enrolled 1,999
Actual Number of Sites Enrolled 99
Patient Follow-up Rate One-thousand and fourteen patients reached 3 years follow-up and attrition is 49.3%. Of the 985 study exits, 450 (46%) were patient deaths.
Final Safety Findings The event-free proportion at 3-years was 89.5% with 425 patients at risk. The 36-month estimate is 89.8% with 555 patients at risk.
Final Effect Findings The proportion of patients with elevated thresholds through 36-months was 15.96% (models 2187/2188/4193/4194) and 8.02% (model 4194).
The proportion of patients with elevated thresholds through 36-months was 15.96% (models 2187/2188/4193/4194) and 8.02% (model 4194).
The proportion of patients with impedance out of range was 0% through 36 months for both lead models.
The proportion of patients with an R-wave out-of-range through 36-months was 1.01% (models 2187/2188/4193/4194) and 0.63% (model 4194).
Study Strengths & Weaknesses One-thousand and fourteen patients reached 3 years follow-up and attrition is 49.3%. Of the 985 study exits, 450 (46%) were patient deaths. The high attrition introduces bias into the study results and limits the external validity.
Recommendations for Labeling Changes Yes, the sponsor will be asked to include a summary of the post-approval study methods and results specifically presenting a frequency histogram of various lead capture outcome values for the InSync registry to be included in the manuals for the applicable devices


Registry Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
PAS Interim Report 02/27/2008 02/27/2008 On Time
7 year report 08/28/2008 08/26/2008 On Time
8 year report 08/28/2009 08/28/2009 On Time
9 year report 08/28/2010 08/24/2010 On Time
10 year report-final report 08/28/2011 06/27/2011 On Time