The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

SJM Brady MRI PAS


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Completed
Application Number /
Requirement Number
P140033 / PAS001
Date Original Protocol Accepted 01/31/2017
Date Current Protocol Accepted 05/22/2020
Study Name SJM Brady MRI PAS
Device Name Assurity MRI and Endurity MRI Pacemakers, Tendril MRI Lead, MRI Activator, Merlin PCS Programer Software
Clinical Trial Number(s) NCT01576016 
General Study Protocol Parameters
Study Design Comprehensive/Linked/RegistryBased Surveillance
Data Source Admin Database
Comparison Group Historical Control
Analysis Type Analytical
Study Population Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives Primary Objective: To determine the Tendril MRI RA and Tendril MRI RV lead-related complication-free rates at 5 years.
Secondary Objective: To determine the rate of cardiac tamponade or cardiac perforation events through 5 years.
The real-world approach to identify lead-related complications will use Abbott-maintained internal databases combined with CMS-maintained Medicare claims data. Every Abbott device has a unique combination of model and serial number that is preserved across Abbott databases and can be used to link data from these databases together.
Study Population All patients who has an implant of an SJM system along with a Tendril MRI RA and/or RV lead, meet inclusion/exclusion criteria, and link to Medicare FFS claims will be included in the analysis of the endpoints.
Sample Size Assuming the time to first complication follows an exponential distribution and the complication free survival at 5 years is 95%, a minimal sample size of 645 Tendril MRI RA and 645 Tendril MRI RV leads provides 80% power to reject the null hypotheses of 92.5% complication-free survival years at the 5% significance level. It is estimated that 23,230 Tendril MRI RA and 22,839 Tendril MRI RV lead implants will meet inclusion /exclusion criteria for this study, which greatly exceeds the minimal required sample size.
Key Study Endpoints Primary Endpoint: Freedom from Tendril MRI RA and/or RV lead-related complications through 5-years of follow-up.
Secondary Endpoint: The secondary outcome measure is the rate of cardiac tamponade or cardiac perforation events. The objective is to characterize the rate of these events. All cardiac tamponade and cardiac perforations will be evaluated for their relationship to the Tendril MRI RA or Tendril MRI RV leads, reported and summarized.
Follow-up Visits and Length of Follow-up The study completion is 5 years after the final implant, June 30, 2018. The data retention period will be for 2 years after the submission of the final report.
Interim or Final Data Summary
Actual Number of Patients Enrolled 25,601
Actual Number of Sites Enrolled N/A
Patient Follow-up Rate N/A
Final Safety Findings • Primary
o Freedom from Tendril MRI lead-related complications through 5 years of follow up
- RA leads: 98.2% (95% Lower Confidence Bound) vs Performance Goal of > 92.5%
- RV leads: 98.3% (95% Lower Confidence Bound) vs Performance Goal of > 92.5%
• Secondary
o Characterization of rate (worst-case) of cardiac tamponade events
- RA leads: 0.7378%
- RV leads: 0.6692%
o Characterization of rate (worst-case) of cardiac perforation events
-RA leads: 0.1042%
-RV leads: 0.0793%
Final Effect Findings N/A
Study Strengths & Weaknesses • Strengths
o The RWE analysis included a much larger study population compared to a traditional prospective study design.
o Fewer patients are lost to follow up in RWE studies since patients are not required to attend follow-up visits.
o The study included a wide sample of operators and institutions.
• Weaknesses
o There is an approximate lag time of 5-8 months for availability of Medicare claims data for analysis.
o RWE analysis is limited by the lack of specificity in various types of lead related complications that were determined based on diagnosis codes in Medicare claims data

Recommendations for Labeling Changes Yes, recommend including a summary of the PAS including results in the IFU manual.


SJM Brady MRI PAS Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
6 month report 08/01/2017 08/01/2017 On Time
1 year report 01/31/2018 01/31/2018 On Time
18 month report 08/01/2018 08/01/2018 On Time
2 year report 01/31/2019 01/29/2019 On Time
3 year report 01/31/2020 01/31/2020 On Time
4 year report 01/30/2021 01/26/2021 On Time
5 year report 01/30/2022 02/09/2022 Overdue/Received
6 year report 01/30/2023 01/30/2023 On Time
final report 01/30/2024 01/31/2024 Overdue/Received