The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

BIOFLEX-I PAS


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Completed
Application Number /
Requirement Number
P160025 / PAS001
Date Original Protocol Accepted 03/23/2017
Date Current Protocol Accepted 08/07/2017
Study Name BIOFLEX-I PAS
Device Name ASTRON PULSAR STENT SYSTEM, PULSAR-18 STENT SYSTEM
Clinical Trial Number(s) NCT01319812 NCT01661231 
General Study Protocol Parameters
Study Design Prospective Cohort Study
Data Source New Data Collection
Comparison Group Historical Control
Analysis Type Analytical
Study Population Transit.Adolescent B(as adults): 18 yrs < 22 yrs, Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives Prospective, non-randomized, multi-center follow-up of the BIOFLEX-I pivotal study (G100002)
Study Population The study population included subjects greater than or equal to 18 years old with life-style limiting claudication or rest pain, with an ABI less than or equal to 0.9 (resting or exercise). Inclusion criteria included one de novo, restenotic or occluded femoropopliteal lesion, OR two de novo , restenotic or occluded lesions representing one femoropopliteal indication and one iliac indication on contralateral limbs (i.e., one lesion per limb). Lesion length was specified as less than or equal to 190 mm (if de novo or restenotic) or less than or equal to 100mm (if occluded), with target vessel reference diameter: 2.5 to 6 mm (SFA/PPA) by visual estimate and with angiographic evidence of atleast one distal vessel runoff to the foot.

Sample Size A total of 280 subjects were active at the time of the 12-month primary endpoint assessment. The long-term post-market analysis of the Pulsar stent group will include a minimum of 202 subjects (238 subjects x 0.85) at 36 months and a minimum of 238 subjects (280 subjects x 0.85) at 24 months. This calculation accounts for a 15% annual attrition rate due to premature exits including subject withdrawal and lost to follow (LTF) and excluding deaths.
Key Study Endpoints The primary endpoint to be assessed is freedom from a composite of clinically-driven target lesion revascularization (TLR) and index limb amputation at 36 months, as defined by the protocol. The secondary endpoints to be assessed include the following:
Freedom from a composite of clinically-driven TLR and index limb amputation at 24 months.
Major adverse events (MAE) rate (30-day mortality, clinically-driven TLR, and index limb amputation) and their components at 24 months.
MAE rate (30-day mortality, clinically driven TLR, and index limb amputation) and their components at 36 months.
Target lesion revascularization (TLR) rate at 24 months.
Target lesion revascularization (TLR) rate at 36 months.
Stent fracture rate at 24 months.
Stent fracture rate at 36 months.
Adverse events rates at 36 months.
Follow-up Visits and Length of Follow-up 36 months
Interim or Final Data Summary
Actual Number of Patients Enrolled 782
Actual Number of Sites Enrolled 43
Patient Follow-up Rate 1, 6, 12, 24, and 36 months
Final Safety Findings Primary- Freedom from clinically-driven target lesion revascularization (TLR) and index limb amputation: 78.2% (201/257), 95% CI [72.7%, 83.1%] at 36 months
Final Effect Findings Primary – primary patency of the Astron Pulsar stent at 12 months: 66.8% (179/268), 95% CI [60.8%, 72.5%]
Study Strengths & Weaknesses NONE
Recommendations for Labeling Changes Yes, update to include PAS data


BIOFLEX-I PAS Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
one year report/final report 03/23/2018 03/23/2018 On Time