The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

MANAGE-HF


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Completed
Application Number /
Requirement Number
P010012 S436/ PAS002
Date Original Protocol Accepted 07/25/2017
Date Current Protocol Accepted 08/26/2022
Study Name MANAGE-HF
Device Name NG4 Family of Pulse Generators CRT-DS
General Study Protocol Parameters
Study Design Prospective Cohort Study
Data Source New Data Collection
Comparison Group Device Subjects Serve as Own Control
Analysis Type Analytical
Detailed Study Protocol Parameters
Study Objectives 1. To evaluate and optimize HeartLogic Heart Failure Diagnostics clinical integration and the alert management process (Phase I).
2. To evaluate Heart Logic in regards to patient outcomes of death and heart failure hospitalizations (Phase II).
Study Population Test Device: Phases I and II: ICD and CRT-D devices with HeartLogic turned ON
Control Device: Phase II: ICD and CRT-D devices with HeartLogic alerts turned OFF
Sample Size Between 1700 and 2700 subjects will be enrolled in the study:
Phase I: Up to 200 subjects
Up to 200 subjects are required in Phase I to build critical knowledge regarding the clinical integration of HeartLogic and the alert management process.
Phase II: 1500-2500 subjects
The sample size required for Phase II of MANAGE-HF is event-driven.
Between 1500 and 2500 subjects will be enrolled in the trial to accrue the necessary number of endpoint events to sufficiently power the primary endpoint. A total of 864 endpoints events from randomized subjects in Phase II are required to sufficiently power the primary endpoint, based on the following assumptions:
Expected control group incidence of annual heart failure
hospitalization and death = 20%
One-sided Alpha = 5%
Power = 90%
Hazard Ratio = 0.8
Estimated enrollment period = 24 months
Expected Follow-up period = minimum of 24 months for last patient enrolled
Final sample size will be determined during the study using a sample size re-estimation interim analysis.
Key Study Endpoints LATITUDE NXT 5.0 System
Follow-up Visits and Length of Follow-up Follow-Up Schedule:
Enrollment Visit
Baseline Visit
Scheduled follow-up Visits
Annual Phone Calls (6, 18,30, 42 months… from baseline)
Annual In-Clinic (12,24,36,48 months…from baseline visit)
Additional Follow-up Visits
HeartLogic Alert Action Visit
Additional Subject Interactions (i.e., Not in Response to HeartLogic Alerts)
End of Study Visit
Interim or Final Data Summary
Actual Number of Patients Enrolled Phase 1: 200 subjects
Phase 2: 3886 subjects
Actual Number of Sites Enrolled Phase 1: 29 sites
Phase 2: N/A
Patient Follow-up Rate Phase 1
Rate = 72%
191 completed the baseline visit
144 followed until the last patient enrolled completed their 12-
month visit
37 withdrawals
19 subjects died

Phase 2:
N=1458 patients were included in the analysis, of which 19 died or their device was no longer in service. These were followed for an average of 1.08 years
Final Safety Findings Phase 1 does not have a safety endpoint
Phase 2 does not have a safety endpoint
Final Effect Findings Phase 1 does not have effectiveness endpoints

Phase 2: The observed Sensitivity was 74.5% and an adjusted Sensitivity (95% confidence interval) was 71.3% (65.4%, 76.6%). The False
Positive rate (95% confidence interval) was 1.48 (1.42, 1.59).

Study Strengths & Weaknesses Phase 1: Patients that were included in Phase I of MANAGE-HF were required to meet specific study inclusion and exclusion criteria. This
included evidence of worsened heart failure prior to enrollment. In addition, patients with severe kidney impairment or short life
expectancy were excluded. Thus, the results are limited to the population that was studied. Furthermore, the small sample size and
lack of a control group limits the interpretation of the results.
Finally, the study was primarily conducted in the US, and results may vary in other geographies.

Phase 2:
• The pre-specified endpoints were met.

• Analysis was retrospective
- The data source for HFE was CMS claims, so there was no event
or site monitoring.
- Events were not independently adjudicated.
- Some results were coarsened due to low numbers as required by
CMS reporting policies.
- Only US patients are included, and results may be different in
other geographies.
- Only CMS patients were included, and on average were older
with more comorbidities than the patients included in the pre-
market (MultiSENSE) study.
Recommendations for Labeling Changes Phase 1: No labeling changes are recommended.
Phase 2: Labeling changes are recommend.


MANAGE-HF Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
six month report 10/31/2017 11/02/2017 Overdue/Received
one year report 05/02/2018 04/30/2018 On Time
18 month report 10/31/2018 10/31/2018 On Time
interim report 12/17/2018 12/17/2018 On Time
two year report 05/02/2019 05/01/2019 On Time
interim report 02/04/2020 02/04/2020 On Time
three year report 05/01/2020 05/01/2020 On Time
4 year report 05/01/2021 04/30/2021 On Time
5 year report 05/01/2022 08/08/2022 Overdue/Received
phase 1 final report 12/07/2022 12/07/2022 On Time
final report 02/28/2023 02/28/2023 On Time