Post-Approval Studies (PAS) Database

  • Print
  • Share
  • E-mail
-

The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

QP ExCELS Sentus QP Ext CRT w/ Quad LV


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Completed
Application Number /
Requirement Number
P070008 S079/ PAS001
Date Original Protocol Accepted 05/04/2017
Date Current Protocol Accepted 09/24/2019
Study Name QP ExCELS Sentus QP Ext CRT w/ Quad LV
Device Name PACEMAKER ICD CRT NON IMPLANTED COMPONENTS
Clinical Trial Number(s) NCT02290028  
General Study Protocol Parameters
Study Design Prospective Cohort Study
Data Source New Data Collection
Comparison Group Objective Performance Criterion
Analysis Type Descriptive
Study Population Transit.Adolescent B(as adults): 18 yrs < 22 yrs, Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives Confirm long-term safety of the Biotronik Sentus OTW QP left ventricular pacing leads. The evaluation of safety will be freedom from Sentus QP related complications at 5 years post-implant.
Study Population CRT-eligible Heart Failure patients
Sample Size Postmarket Sample Size Required: 979
Postmarket Sample Size Required including attrition = 1754
Key Study Endpoints Primary Endpoint: The purpose is to evaluate the long-term Sentus related complication-free rate through 5 years post-implant.
Follow-up Visits and Length of Follow-up 5 years
Interim or Final Data Summary
Actual Number of Patients Enrolled 2872
Actual Number of Sites Enrolled N/A - RWE
Patient Follow-up Rate N/A - RWE
Final Safety Findings Primary safety objective of complication-free rate (CFR) greater than 92.5% at 5 years post-implant for Sentus LV leads was met. 84 complications noted in 2920 Sentus leads. Lower bound of the two-sided 95% confidence interval of CFR was 92.72%.
Final Effect Findings N/A
Study Strengths & Weaknesses This study’s use of real-world data captures a large representative assessment of device performance. While data granularity is limited, the study confirms there are no significant safety concerns with the device’s long term performance.
Recommendations for Labeling Changes Labeling changes should include the study design, results, and strengths and weaknesses of the PAS.


QP ExCELS Sentus QP Ext CRT w/ Quad LV Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
6 month report 11/02/2017 11/02/2017 On Time
1 year report 11/04/2018 11/02/2018 On Time
2 year report 11/04/2019 10/31/2019 On Time
3 year report 11/04/2020 11/02/2020 On Time
4 year report 11/04/2021 11/03/2021 On Time
5 year report 11/04/2022 11/02/2022 On Time
6 year report 11/04/2023 10/31/2023 On Time
final report 11/04/2024 10/30/2024 On Time