The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

Continued f/u of Premarket Pivotal Cohort


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Completed
Application Number /
Requirement Number
P160054 S008/ PAS001
Date Original Protocol Accepted 01/18/2019
Date Current Protocol Accepted  
Study Name Continued f/u of Premarket Pivotal Cohort
Device Name HeartMate 3 Left Ventricular Assist System
Clinical Trial Number(s) NCT02224755  
General Study Protocol Parameters
Study Design Prospective Cohort Study
Data Source New Data Collection
Comparison Group Concurrent Control
Analysis Type Descriptive
Study Population Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives To characterize the clinical outcomes through 5 years post implantation by continuing to follow the premarket pivotal cohort of patients.
Study Population Subjects (both HeartMate 3 and HeartMate II) who were enrolled in the premarket pivotal cohort.
Sample Size All living subjects (about 533).
Key Study Endpoints The composite endpoint of survival to transplant or recovery, or
on left ventricular assist device (LVAD) support free of debilitating stroke (Modified Rankin Score > 3) or reoperation to replace the pump; mortality; bleeding; major infection; hemolysis; device thrombosis; neurological dysfunction; and any other serious adverse events; as well as New York Heart Association (NYHA) classification and 6-minute walk distance (6MWD).
Follow-up Visits and Length of Follow-up 5 years
Interim or Final Data Summary
Actual Number of Patients Enrolled 295 subjects enrolled
Actual Number of Sites Enrolled 53
Patient Follow-up Rate The follow-up rate through 5 years was 98.67%.
Final Safety Findings Survival Free of Debilitating Stroke or Reoperation at 5 years The Kaplan- Meier estimate of survival free of debilitating stroke or
reoperation to replace the pump was higher for the HM3 LVAS (69.7%) compared to the HMII LVAS (41.9%).

Survival and Subject Outcomes at 5 years
The number of at risk subjects is reduced for both HM3 (134) and HMII (82) subjects. The percent survived for HM3 is 73.7% and HMII 55.6%.
The sponsor reports subjects with HM3 LVAD have a statistically significant higher, p=0.0001, survival rate than subjects with HMII.

Subgroup Analyses
There was no difference in outcomes based on gender. The sponsor reports male and female subjects who received HM3 had statistically
significant higher, p<0.0001, survival free of debilitating stroke or reoperation than subjects receiving HMII. In addition the sponsor
reports for INTERMACS 1-2 and 3-7 profiles, subjects who received HM3 had statistically significant higher survival (p=0.0223
INTERMACS 1-2, p<0.0001 INTERMACS 3-7) free of debilitating stroke or reoperation than subjects receiving HMII.

Device related SAEs
The sponsor reports HM3 had significantly less bleeding (p<0.001), stroke (p=0.035) and confirmed device thrombosis events (p<0.001)
than HMII subjects. Although, the sponsor reports significantly less bleeding with the HM3, bleeding and major infection are still a risk.
Major Infection HMII: # of subjects 34; % of subjects 29.1%; # of events 67; EPPY 0.228, HM3: # of subjects 70; % of subjects 39.3%; #
of events 122; EPPY 0.263, p- value 0.348 Bleeding HMII: # of subjects 38; % of subjects 32.5%; # of events 124;
EPPY 0.4228, HM3: # of subjects 50; % of subjects 28.1%; # of events 99; EPPY 0.213, p- value <0.001
Final Effect Findings The NYHA classification and six-minute walk distance assessments were not completed consistently at follow-up visits. Limited data
points and comparisons can be made. Overall, both HM3 and HMII subjects had improvements in NYHA classification and 6MWD
assessment throughout the 5 year follow-up compared to baseline.
Study Strengths & Weaknesses This extended follow-up study provides continued evaluation through five years post implant on the safety and effectiveness of the HM3
LVAS. The study provides additional data on the composite endpoint as well as mortality, bleeding, major infection, hemolysis, device
thrombosis, neurological dysfunction, device related serious adverse events and functional status. Of note, there was delayed site activation,
sites declined to participate, not all subjects reconsented from the MOMENTUM 3 IDE and the COVID-19 pandemic, which attributed to
data not being collected and impact the results.
Recommendations for Labeling Changes A labeling update will be requested to add five-year performance data to inform clinicians and patients of benefit risk profile for the device


Continued f/u of Premarket Pivotal Cohort Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
1 year report 10/18/2019 10/17/2019 On Time
2 year report 10/17/2020 10/15/2020 On Time
3 year report 10/17/2021 10/18/2021 Overdue/Received
final report 10/17/2022 06/02/2022 On Time