The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

ZEVR-Zephyr Valve Registry


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Ongoing
Application Number /
Requirement Number
P180002 S005/ PAS001
Date Original Protocol Accepted 10/02/2019
Date Current Protocol Accepted 03/25/2023
Study Name ZEVR-Zephyr Valve Registry
Device Name Zephyr Endobronchial Valve (EBV) System
General Study Protocol Parameters
Study Design Prospective Cohort Study
Data Source Sponsor Registry
Comparison Group Concurrent Control
Analysis Type Descriptive
Study Population Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives Prospective Cohort Study: Multi-center, single-arm, prospective Registry to be conducted at up to 15 clinical sites. Approximately 150 patients with severe emphysema undergoing Zephyr Valve treatment will be enrolled and followed out to 3 years. Assessments will be performed at 45-days, 6-months, 12-months, 24-months and 36-months post-procedure. The Zephyr Valve Registry will be conducted at a minimum of 5, and a maximum of 10 clinical sites. Enrollment is anticipated to be completed over 24 months. A minimum of 20 subjects treated with at least one Zephyr 5.5-LP EBV will be enrolled.
Study Population Adult patients with hyperinflation associated with severe emphysema in regions of the lung that have little to no collateral ventilation. Concurrent control comparison group.
Sample Size 150
Key Study Endpoints Safety endpoints:
Incidence rate of Pneumothorax at 45-days post-procedure.
Incidence rate of device/procedure related Serious Adverse Events (SAE) including but not limited to COPD exacerbations, pneumonia, hemoptysis, valve expectoration, valve migration and respiratory failure.
Effectiveness endpoints:
Treated Lobar Volume Reduction (TLVR) assessed by HRCT at 45-days.
Forced Expiratory Volume in 1 second (FEV1) at 6-months, 12-months, 24-months and 36-months.
6MWD at 6-months and 12-months.
SGRQ at 6-months and 12-months.
BODE Index at 6-months and 12-months.
Percent of patients with a TLVR of =350mL10,11.
Percent of patients with a TLVR of =563mL12.
Responder rates reported as percent of patients meeting the Minimal Clinically Important Difference (MCID) for each variable at 6-months and 12-months.
- MCID for FEV1: Increase of 10% (Liters)
- MCID for 6MWD: Increase of +25 meters
- MCID for SGRQ: Decrease of 4 points

Follow-up Visits and Length of Follow-up 36 months
Interim or Final Data Summary
Interim Results Safety Results
Pneumothorax adverse event rate: 39.5%

Effectiveness Results
The mean reduction in TLV at 45 days was 1185.3 mL and at 12-months was 1185.0 mL representing a percent reduction of 64.37% and 65.47%, respectively.

Actual Number of Patients Enrolled 162
Actual Number of Sites Enrolled 12
Patient Follow-up Rate 1-year post-procedure follow up: 68.5%


ZEVR-Zephyr Valve Registry Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
6 month report 12/28/2019 12/18/2019 On Time
1 year report 06/28/2020 06/25/2020 On Time
2 year report 06/28/2021 06/28/2021 On Time
3 year report 06/28/2022 06/28/2022 On Time
4 year report 06/28/2023 06/22/2023 On Time
5 year report 06/28/2024 06/26/2024 On Time
6 year report 06/28/2025 07/02/2025 Overdue/Received
7 year report 06/28/2026 06/26/2026 On Time
8 year report 06/28/2027    