| General |
| Study Status |
Ongoing |
Application Number / Requirement Number |
P140031 S085/ PAS001 |
| Date Original Protocol Accepted |
08/16/2019
|
| Date Current Protocol Accepted |
 
|
| Study Name |
Cont f/u of SAPIEN 3 Low Risk cohort PAS
|
| Device Name |
SAPIEN 3 Ultra transcatheter heart valve
|
| Clinical Trial Number(s) |
NCT02675114
|
| General Study Protocol Parameters |
| Study Design |
Prospective Cohort Study
|
| Data Source |
New Data Collection
|
| Comparison Group |
Concurrent Control
|
| Analysis Type |
Descriptive
|
| Study Population |
Adult: At least 22 yrs
|
| Detailed Study Protocol Parameters |
| Study Objectives |
Continued follow-up of all living subjects who were enrolled in the pivotal cohort under the IDE. The objective of this study is to characterize the clinical outcomes annually through 10 years post-procedure.
|
| Study Population |
All living subjects who were enrolled in the IDE pivotal cohort.
|
| Sample Size |
All living subjects.
|
| Key Study Endpoints |
All-cause mortality, all stroke (disabling and non-disabling), life-threatening bleeding, requirement for renal replacement, coronary artery obstruction requiring intervention, major vascular complication, valve-related dysfunction requiring repeat procedure, new permanent pacemaker implantation, prosthetic valve endocarditis, prosthetic valve thrombosis, New York Heart Association (NYHA) classification, 6-Minute Walk Test (6MWT) distance, Kansas City Cardiomyopathy Questionnaire (KCCQ) score, and hemodynamic performance metrics by Doppler echocardiography.
|
| Follow-up Visits and Length of Follow-up |
10 years
|
| Interim or Final Data Summary |
| Interim Results |
At 8 years: All-cause mortality (including vital status check): 26.5% [22.8, 30.8]* (TAVR); 22.2% [18.5, 26.5]* (SAVR) - KM Stroke: 9.4% [7.0, 12.6]* (TAVR); 8.2% [5.9, 11.4]* (SAVR) - KM Aortic valve reintervention: 7.3% [5.2, 9.9]* (TAVR); 6.8% [4.6, 9.6]* (SAVR) - CIF NYHA (% in class I/II): 95.1% (TAVR); 95.4% (SAVR) KCCQ (Mean Change from Baseline): 12.3 (TAVR); 13.9 (SAVR) At 7 years: Bioprosthetic valve failure: 6.9% [4.8, 9.4]* (TAVR); 7.5% (SAVR) [5.2, 10.4]* - CIF Transvalvular mean gradient: 13.3 (TAVR); 12.1 (SAVR) Effective orifice area: 1.9 (TAVR); 1.8 (SAVR) Greater than or equal to moderate total aortic regurgitation: 2.4% (TAVR); 1.6% (SAVR) Greater than or equal to moderate PVL: 1.0% (TAVR); 0.4% (SAVR) *Confidence Interval
|
| Actual Number of Patients Enrolled |
1,000 randomized subjects; 948 implanted (495 TAVR & 453 SAVR)
|
| Actual Number of Sites Enrolled |
71
|
| Patient Follow-up Rate |
At 8 years (patient status known): 89.7% (TAVR); 80.4% (SAVR)
|