The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

Extended Follow-up of IDE Subjects


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Completed
Application Number /
Requirement Number
P200022 / PAS001
Date Original Protocol Accepted 11/05/2020
Date Current Protocol Accepted 05/26/2021
Study Name Extended Follow-up of IDE Subjects
Device Name Simplify® Cervical Artificial Disc
Clinical Trial Number(s) NCT02667067 
General Study Protocol Parameters
Study Design Prospective Cohort Study
Data Source New Data Collection
Comparison Group Historical Control
Analysis Type Analytical
Study Population Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives Objective: To evaluate the 5-year long-term safety and effectiveness of the Simplify Cervical Artificial Disc in subjects with symptomatic cervical
radiculopathy with or without neck pain or myelopathy due to a single level abnormality localized to the level of the disc space that is
unresponsive to conservative management.

Design: A continuation of the prospective, multi-center, historically controlled IDE study population to evaluate the long-term safety and
effectiveness of the Simplify Cervical Artificial Disc.
Study Population Investigational Device: Simplify Cervical Artificial Disc
Historical Control: Anterior Cervical Discectomy and Fusion (ACDF)
Sample Size 142 Simplify Cervical Artificial Disc IDE subjects and 106 historical ACDF control subjects are elgible for for the PAS study
Key Study Endpoints Safety Endpoints:
• Collection and evaluation of all adverse events for severity and device and/or procedure-relatedness
• Evaluation of device failure (device breakage, migration, or mechanical failure) as defined in Radiographic Evaluation Protocol
• Subsequent surgical procedures (revision, removal, reoperation, or supplemental fixation) at the index level
• Neurologic evaluation – investigational device only
• Dysphagia Handicap Index (DHI) – investigational device only
Effectiveness Endpoints:
• Neck Disability Index (NDI) assessment
• Radiographic evaluation
• VAS pain assessments (neck and arm; neck, left arm and right arm (investigational device only))
• Treatment satisfaction
• SF-12 Health Survey
Follow-up Visits and Length of Follow-up 5 years
Interim or Final Data Summary
Actual Number of Patients Enrolled 132 investigational device subjects, 106 historical ACDF control subjects enrolled in PAS.
Actual Number of Sites Enrolled 15 of 17 investigational device sites, 21 of 24 historical control sites
Patient Follow-up Rate At month 60, the follow-up rate is 89% for the investigational device and 61% for the PS selected ACDF control subjects.
Final Safety Findings In this PAS, the overall AE and SAE rates were comparable between treatment groups. There was no difference between groups with regard to AEs categorized as definitely device- or procedure-related; however, there was a slightly higher rate of device-related spasm AEs in the Simplify group and a slightly higher rate of device-realted pain (narcotic given) in the ACDF control group. There were two instances of secondary surgical interventions (SSIs) in the Simplify group and five SSIs in the ACDF control group after the 24-month timepoint. See tables below.
Final Effect Findings The composite clinical success for the Simplify Disc group was 91.1% compared to 74.6% for the historical ACDF control group. The study success criteria for non-inferiority as well as for superiority have been met.
Study Strengths & Weaknesses This 5-year follow-up study had high follow-up rates for the investigational group but lower than the recommended 85% follow-up rate for the historical control group.
Recommendations for Labeling Changes It is recommended that the labeling be updated with the results of the PAS study (i.e., data at 60-months).


Extended Follow-up of IDE Subjects Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
1 year report 09/18/2021 09/20/2021 Overdue/Received
2 year report 09/18/2022 09/16/2022 Overdue
final report 09/18/2023 07/24/2023 On Time