• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Post-Approval Studies (PAS) Database

  • Print
  • Share
  • E-mail
-

The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

SAPIEN 3 and SAPIEN 3 Ultra Aortic THV-in-THV Surv


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Ongoing
Application Number /
Requirement Number
P140031 S112/ PAS001
Date Original Protocol Accepted 09/09/2020
Date Current Protocol Accepted 09/09/2020
Study Name SAPIEN 3 and SAPIEN 3 Ultra Aortic THV-in-THV Surv
Device Name EDWARDS SAPIEN 3 and SAPIEN 3 Ultra Transcatheter Heart Valve System
General Study Protocol Parameters
Study Design Active Surveillance
Data Source External Registry
Comparison Group Historical Control
Analysis Type Descriptive
Detailed Study Protocol Parameters
Study Objectives Characterize long-term clinical outcomes for the SAPIEN 3 and SAPIEN 3 Ultra THV Systems in patients with a failing aortic transcatheter bioprosthetic valve after PMA approval through the Society of Thoracic Surgeons (STS)/American College of Cardiology (ACC) Transcatheter Valve Therapy (TVT) registry-based surveillance.
Study Population All patients entered into the TVT Registry database and treated with the SAPIEN 3 or SAPIEN 3 Ultra System for the first 3 years after PMA approval will be included as part of this registry-based post-approval surveillance.
Sample Size All consecutive patients treated within the first 3 years after the PMA approval that are entered into the TVT Registry (enrollment period).
Key Study Endpoints For aortic THV-in-THV patients, the surveillance is monitoring the following outcomes:
Procedural: Device implanted successfully
30 Days: Myocardial infarction
30 Days, 12 months, and annually for 10 years: all-cause mortality, all stroke, and valve reintervention
30 Days & 1 year: TIA, life-threatening/major bleeding, vascular complications, new requirement for dialysis, KCCQ, mean gradient, and paravalvular leak
Follow-up Visits and Length of Follow-up 10 years
Interim or Final Data Summary
Interim Results Safety Endpoints:
All-cause death: 4.5% (30 Days); 18.9% (1Y)
Cardiac death: 2.6% (30 Days); 7.6% (1Y)
All strokes: 1.7% (30 Days); 2.5% (1Y)
Aortic Valve reintervention: 0.2% (30 Days); 0.2% (1Y)
Life threatening/major bleeding: 1.5% (30 Days); 4.2% (1Y)
Myocardial infarction: 0.5% (30 Days); N/A (1Y)
New requirement for dialysis: 0.5% (30 Days); 0.5% (1Y)
TIA: 0.2% (30 Days); 0.2% (1Y)
Vascular Complications: 4.0% (30 Days); 4.0% (1Y)
Major vascular complications: 0.7% (30 Days); 0.7% (1Y)
Minor vascular complication: 3.3% (30 Days); 3.3% (1Y)
Paravalvular Leak: 30-days: >= Moderate – 1.4% (4/283) 1-year: >= Moderate – 0.8% (1/125)
KCCQ: Baseline: 37.1; 30-day: 73.9; 1-year: 78.6;
Mean Gradient: Baseline: 39.8mmHg; 30-day: 16.4mmHg; 1-year: 15.5mmHg
Effectiveness Endpoints:
Paravalvular Leak:
30-days: > Moderate – 1.4% (4/283)
1-year: > Moderate – 0.6% (1/168)
KCCQ: Baseline: 37.0; 30-day: 73.7; 1-year: 79.4; Mean Gradient: Baseline: 39.7mmHg; 30-day: 16.4mmHg; 1-year: 15.5mmHg
Actual Number of Patients Enrolled 434 attempted; 432 implanted with the valve
Actual Number of Sites Enrolled 216
Patient Follow-up Rate 30 Days: 96.1% (417/434); 1 year: 81.1% (352/434)
30 Days: 5 lost to follow-up; 12 missed visits
1 year: 17 lost to follow-up; 65 missed visits


SAPIEN 3 and SAPIEN 3 Ultra Aortic THV-in-THV Surv Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
1 year report 09/09/2021 09/07/2021 On Time
2 year report 09/09/2022 08/30/2022 On Time
3 year report 09/09/2023 09/11/2023 Overdue/Received
4 year report 09/08/2024 09/03/2024 On Time
5 year report 09/08/2025 08/28/2025 On Time
6 year report 09/08/2026 09/02/2026 On Time
7 year report 09/08/2027    


Contact Us

Mandated Studies Program
Food and Drug Administration
10903 New Hampshire Ave.
Silver Spring, MD 20993-0002
Email: MandatedStudiesPrograms@fda.hhs.gov

Additional Resources

-
-