The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

Continued Follow-Up of G170301 Clinical Study


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Completed
Application Number /
Requirement Number
P200036 / PAS001
Date Original Protocol Accepted 03/01/2022
Date Current Protocol Accepted 07/15/2022
Study Name Continued Follow-Up of G170301 Clinical Study
Device Name eCoin Peripheral Neurostimulator
Clinical Trial Number(s) NCT03556891 
General Study Protocol Parameters
Study Design Active Surveillance
Data Source Other Data Source
Comparison Group Device Subjects Serve as Own Control
Analysis Type Descriptive
Study Population Transit.Adolescent B(as adults): 18 yrs < 22 yrs, Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives Continued follow-up of premarket cohort to obtain longer term safety and effectiveness data on the eCoin out to 5 years post-implantation
Study Population Subjects diagnosed with Urgency Urinary Incontinence (UUI).
Sample Size Available sample size for longer term follow-up of the original cohort.
The original IDE study was designed to enroll and implant between 120 and 135 patients; it enrolled 137 patients with 4 patients exiting prior to implantation (due to ineligibility). Thus, 133 patients were implanted.
Key Study Endpoints To collect and report clinical outcomes for safety and effectiveness, including 3-day bladder diaries, all device- or procedure-related adverse events, including but not limited to, infections, vasculitis, cellulitis, neuro-sensory and -motor events, device migrations, erosions, unplanned explantations, and re-implantation, to FDA through 5 years post-implantation on patients enrolled in G170301.

Follow-up Visits and Length of Follow-up 5 years, post-implantation
Interim or Final Data Summary
Actual Number of Patients Enrolled 31
Actual Number of Sites Enrolled 9
Patient Follow-up Rate 67.7% (21 of 31 at 5 years)
The study reconsented 31 subjects for continued follow-up out to 5 years post-implantation. Of those reconsented, 15 subjects consented to re-enroll for follow-up only (no device re-implantation) and 16 subjects consented to be explanted and re-implanted. Prior to the 205 week follow-up, one subject withdrew and 7 subjects were missing at random. Prior to the 257 week follow-up, 2 additional subjects were missing at random.
Final Safety Findings There were no unanticipated adverse device events. Overall, there were 2 device related adverse events: 1 for incisional site pain and 1 procedure-related adverse event of post-operative infection. These events are anticipated for this type of device.
Final Effect Findings For subjects who underwent follow up only and were not re-implanted, 75% (9/12 subjects) showed at least a 50% reduction in UUI episodes at 205 weeks, and 58% (7/12 subjects) of subjects experienced at least a 50% reduction in UUI episodes at 257 weeks. Of subjects who were reimplanted, 45% (5/11 subjects) of subjects experienced at least a 50% reduction in UUI episodes at 205 weeks post activation, and 78% (7/9 subjects) experienced at least a 50% reduction in UUI episodes at 257 weeks.
Study Strengths & Weaknesses Study shows similar results in terms of safety and effectiveness at 4- and 5-years post-implant compared to the pre-market results. However, the results are based upon a very limited number of patients since many subjects did not reconsent and/or were lost to follow-up. Further, the 4- and 5-year data on re-implanted subjects is difficult to interpret as the time elapsed between the explant and re-implant varied among those subjects. Reimplantation of a new eCoin device did not reset their follow-up clock
Recommendations for Labeling Changes None


Continued Follow-Up of G170301 Clinical Study Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
6 month report 08/30/2022 08/31/2022 Overdue/Received
1 year report 03/01/2023 02/24/2023 On Time
18 month report 08/30/2023 08/24/2023 On Time
2 year report 02/29/2024 02/26/2024 On Time
final report 03/01/2025 02/28/2025 On Time