The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

KEYTRUDA PAS


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Completed
Application Number /
Requirement Number
P210001 S001/ PAS001
Date Original Protocol Accepted 05/20/2022
Date Current Protocol Accepted  
Study Name KEYTRUDA PAS
Device Name VENTANA MMR RxDx Panel
Clinical Trial Number(s) NCT02460198   NCT02628067  
General Study Protocol Parameters
Study Design Prospective & Retrospective Study
Data Source Sponsor Registry
Comparison Group Historical Control
Analysis Type Descriptive
Study Population Transit.Adolescent B(as adults): 18 yrs < 22 yrs, Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives The purpose of this study is to confirm in a post marketing setting the clinical effectiveness of the MMR RxDx as a companion diagnostic in identifying dMMR solid tumor patients for treatment with pembrolizumab.
The primary objective of this study is to estimate the objective response rate (ORR) for pembrolizumab treatment among clinical study subjects whose archived solid tumor specimens are dMMR as determined by retrospective testing with the MMR RxDx (CDx).
The study design includes two cohorts of patients with solid tumors treated with pembrolizumab:
- 34 patients from the original approval cohort (KEYNOTE-158, Cohort K) which previously did not have a full 26-weeks of follow-up, were not included in the original data set, had archival tissue available for CDx testing
- Up to 30 patients prospectively enrolled in the ongoing Cohort L underway in China that have been tested with the CDx
Study Population The study consists of 34 participants from KN-158 Cohort K and 30 participants from KN-158 Cohort K in KN158 is an ongoing multicenter, global, open-label, single-arm cohort of pembrolizumab in participants with any advanced MSI-H/dMMR solid tumor (except CRC) that was incurable and for which prior standard first-line treatment had failed. Participants were eligible for enrollment into Cohort K based on local testing for dMMR/MSI-H status (referred to as the clinical trial assay (CTA).
The sPMA dataset (P210001/S001, approved on March 21, 2022) included data from 321 KEYNOTE-158 Cohort K participants with at least 26 weeks of follow-up by the data cutoff date on October 5, 2020. Since then, an additional 34 participants enrolled in that cohort now have 26 weeks of follow-up. Data from these 34 participants are part of this proposed study. Samples from these participants have been tested with the CDx. The proposed study will evaluate the response to pembrolizumab treatment in these 34 participants based on their CDx MMR status.
The tumor types represented in these 34 participants include anal (1), breast (2), cervical (1), endometrial, gastric (9), glioma (4), ovarian (1), and small intestine (1). Among these 34 participants, enrolled based on dMMR/MSI-H status per the CTA, whose samples were tested by the CDx, 26 had CDx-positive status and 8 had CDx-negative status.
Cohort L in KN158 is an ongoing study cohort enrolling 30 participants who are from mainland China and are of Chinese descent with any advanced solid tumor (including CRC) that is dMMR/MSI-H determined by the CDx (VENTANA MMR RxDx) and Promega PCR, respectively. KN-158 Cohort L is still enrolling, and the tumor types will be included in the report.
Sample Size 34 patients from KEYNOTE-158 Cohort K will be included, 26 of which are CDx positive and 8 are CDx negative. The sample size is limited by the subjects available for inclusion.
Up to 30 additional patients from KEYNOTE-158 Cohort L will be evaluated. The sample size is limited by the subjects available for inclusion.
Key Study Endpoints The primary study endpoint is the objective response rate (ORR) based on partial or complete response among patients whose tumors are dMMR by the CDx.
Duration of response (DOR) will be reported as an additional endpoint. For Cohort K, the ORR observed in CDx-negative patients will also be reported
Follow-up Visits and Length of Follow-up Patients with 26-weeks of follow-up are included in the study.
Interim or Final Data Summary
Actual Number of Patients Enrolled 34 patients in Cohort K, 30 patients in Cohort L
Actual Number of Sites Enrolled Cohort K: 20 sites across 11 countries; Cohort L: 18 sites in China
Patient Follow-up Rate 100% in Cohort K and Cohort L
Final Safety Findings No new safety findings were observed.
Final Effect Findings For the 34 additional Cohort K patients, 26 pooled CDx-positive patients were reported to have an ORR of 50% (95% CI: 29.9, 70.1%), and the 8 pooled CDx-negative patients were reported to have an ORR of 12.5% (95% CI: 0.3-52.7%). For the 30 Cohort L patients, 21 CDx-positive patients were reported to have an ORR of 66.7% (95% CI: 43.0, 85.4%).
Study Strengths & Weaknesses N/A
Recommendations for Labeling Changes N/A


KEYTRUDA PAS Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
1 year report 03/21/2023 03/20/2023 On Time
18 month report 09/19/2023 09/25/2023 Overdue/Received
final report 03/20/2024 03/19/2024 On Time