The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

Lenstec SBL-3 PAS


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Completed
Application Number /
Requirement Number
P200020 / PAS001
Date Original Protocol Accepted 09/16/2022
Date Current Protocol Accepted 11/17/2022
Study Name Lenstec SBL-3 PAS
Device Name SBL-3 Multifocal Intraocular Lens
Clinical Trial Number(s) NCT02487160 
General Study Protocol Parameters
Study Design Prospective Cohort Study
Data Source New Data Collection
Comparison Group Concurrent Control
Analysis Type Descriptive
Study Population Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives The objective of this post approval study (PAS) is to verify the safety of the SBL-3 multifocal IOL and to determine the risk factors that may be associated with the key study endpoint events (MIOL)(P200020) by comparing it to another approved multifocal IOL. The study will be a prospective, randomized, parallel group design. It will compare the SBL-3 MIOL to the Alcon PanOptix® Trifocal Intraocular Lens Model TFAT00 or Model CCWTT0
The study endpoints (discussed below) will be evaluated for each group and a comparison made at the appropriate time points.

Subjects enrolled will be newly enrolled. Enrollment will be done so in a 2:1 manner, in which twice as many SBL-3 IOLs subjects will be enrolled as the control MIOL.

Enrollment will occur in a single phase (Phase 4).

Study Population Subjects enrolled will be those that are in good general and ocular health, with age related cataracts requiring extraction and replacement with an IOL.

Subjects are eligible for the study if they meet the criteria listed below.
greater than or equal to 22 years of age, of any race and either gender
Operable, age related cataract grade in both eyes
Patients who require an IOL power in the range of 15 D – 30 D only.
Able to comprehend and sign a statement of informed consent
Planned cataract removal by phacoemulsification
Potential postoperative visual acuity of 20/30 Snellen acuity or better in both eyes
In good general and ocular health
Patients with preoperative astigmatism less than or equal to 1.0 D
Clear intraocular media other than cataract in study eyes
The subject must be able to undergo second eye surgery between 7 days and 30 days of the first eye surgery.
Able to competently complete testing
Willing and able to attend study visits

The key exclusion criteria will be:
Previous intraocular surgery
Preoperative distance photopic pupil size of < 3.00 mm
Previous corneal refractive surgery
Any inflammation or edema (swelling) of the cornea
Pterygium with corneal involvement or has the potential of corneal involvement (in the opinion of the Investigator) during the course of the study
Subjects with diagnosed degenerative visual disorders (e.g., macular degeneration or other retinal disorders) that are predicted to cause future acuity losses to a level worse than 20/30 Snellen acuity
Subjects who may reasonably be expected to require a secondary surgical intervention at any time during the study (other than YAG capsulotomy)
Amblyopia
Clinically significant ptosis
Clinically severe corneal dystrophy (e.g., epithelial, stromal, or endothelial dystrophy), keratitis, keratoconjunctivitis, keratouveitis, keratopathy, or kerectasia
Diabetic Retinopathy
Extremely shallow anterior chamber, not due to swollen cataract
Microphthalmia
Previous retinal detachment
Previous corneal transplant
Severe dry eye
Recurrent severe anterior or posterior segment inflammation of unknown etiology
Rubella or traumatic cataract
Iris neovascularization
Glaucoma (medically controlled or uncontrolled)
Additionally, subjects with required additional procedures during surgery, forced mechanical or surgical manipulation of the pupil, other than in the bag implantation of the IOL, or other surgical complications that could confound the results of the study or put the subject at risk of ocular system damage or injury will be discontinued from the study during the surgery.


Sample Size No. of subjects: The test group will enroll up to 330 subjects in order to obtain 300 at the final evaluation. The control group will enroll up to 170 subjects in order to obtain 150 subjects at the final evaluation.
No. of sites: 20
Sites location: US

Key Study Endpoints Safety Endpoints
Primary:
The primary safety endpoints are as follows:

The rates of all implanted eyes (primary eyes + fellow eyes) with secondary surgical interventions (SSI), cumulative over the course of the 6- month study, related to visual symptoms (e.g., glare, halo or starburst or blurred vision) or refractive error. This includes stratified analyses of:

i. Secondary surgical interventions related to visual symptoms or any sort, e.g., glare, halo, starbursts, or blurred vision, (excluding those related to refractive error)
ii. Secondary surgical interventions related to refractive error

2) Rate of all implanted eyes (primary eyes + fellow eyes) with absolute manifest refraction spherical equivalent (MRSE) greater than or equal to 1 diopter (D) from the intended target starting at any visit at 21 days postoperatively, or required SSI related to refractive error at any time in the study (cumulative over the 6-month study)
3) Rate of all implanted eyes (primary eyes + fellow eyes) with changes between any two postoperative visits of absolute MRSE greater than or equal to 1 D, starting at any visit at 21 days postoperatively (cumulative over the 6-month study)
4) Rate of all implanted eyes (primary eyes + fellow eyes) with changes between any two postoperative visits monocular uncorrected distance visual acuity (UCDVA) greater than or equal to 10 letters (plus or minus) starting at any visit at 21 days post-operatively (cumulative over the 6-month study)
5) Rate of subjects with significant difficulty due to variations in distance vision on a questionnaire (given at every visit, including unscheduled, starting at any visit at 21 days postoperatively over the 6-month study) defined as a "significant (Grade of 5 for either question #1 or question #2*) variation in vision.”
*Collected at Questions 1 and 2 of the excerpted NEI RQL-426).
6) Rate of all implanted eyes (primary eyes + fellow eyes) with UCDVA worse than 20/40 at any single visit starting at the 3A visit or later.

Secondary:
Rate of all implanted eyes (primary eyes + fellow eyes) with other types of serious adverse events (ISO 11979-7 historical grid Table E.2 - Posterior chamber IOL adverse event rates)
Non-serious adverse events by comparing the subject incidence of each type of non-serious event between two groups with a Fisher’s exact test.
Follow-up Visits and Length of Follow-up The scheduled follow up visits will be 1-day post-operative, 1-week post-operative, 1-month post-operative, 3-month post-operative and 6-month post-operative. The 6-month post-operative visit will have a scheduled window of +/-3 weeks from the initial surgical date.
The study will run a total of 6 months post-operative.

Interim or Final Data Summary
Actual Number of Patients Enrolled 490 total subjects enrolled,
• 320 subjects were enrolled and 319 were bilaterally implanted with the ClearView 3 MIOL
• 170 subjects were enrolled and 169 were bilaterally implanted with the control (Alcon PanOptix Trifocal IOL).
Actual Number of Sites Enrolled 14
Patient Follow-up Rate 95% (446/490)
Final Safety Findings The ClearView 3 demonstrated no statistical differences of secondary surgical interventions (SSI), SSI related to visual symptoms or SSI related to refractive errors. Additionally, there were no statistical differences between manifest refraction spherical equivalent (MRSE) greater than or equal to 1 diopter (D) from the intended target at the study endpoints between the ClearView 3 and the PanOptix. The rate of all implanted eyes with changes of monocular uncorrected distance visual acuity (UCDVA) greater than or equal to 10 letters (plus or minus) did not show any statistical difference between the ClearView 3 and the PanOptix groups. When given the NEI (National Eye Institute) questionnaire asking subjects to report “How much difficulty do you have seeing because of changes in the clarity of your vision over the course of the day?” and “How often are you bothered by changes in the clarity of your vision over the course of the day?”, the ClearView3 showed no statistical difference to the PanOptix control lens. Furthermore, for rates of UCDVA worse than 0.3 LogMAR, the ClearView 3 did not show statistical difference to the PanOptix IOL but the ClearView 3 group did have a lower rate of subjects that experienced worse than 0.3 LogMAR than the PanOptix.
Lenstec successfully met each primary endpoint with the exception of endpoint 3; Rate of all implanted eyes (primary eyes + fellow eyes) with changes between any two postoperative visits of absolute MRSE =1 D, starting at any visit 21 days post-operatively or later (cumulative over the 6-month study). The outcomes were significantly different
Final Effect Findings N/A
Study Strengths & Weaknesses The study’s design included both strengths and weaknesses. The goal of any clinical trial is to produce reliable and valid data while minimizing bias. Strengths of this study include the prospective, randomized nature of enrollment, including a concurrent control. This ensures that the same population of potential subjects are included in the trial and their data is captured as it is available. The study also had a very high rate of subjects completing the study (>95%), which ensures that those subjects enrolled have their data collected, as it relates to the study endpoints. From a demographic standpoint, there were no differences between the subjects in both groups, which also makes group to group comparison valid.
Potential study weaknesses include non-masked participation of its participants. In this study, patients masking. Masking surgical staff would have been very difficult as the two study articles have considerably different optic and haptic designs. This would be noticed from the time of implantation (surgeon, surgical staff) through to the final study visit (surgeon, clinical staff) who necessarily have to visualize the IOL for the purpose of data capture. Another weakness includes the fact that the amount of experience surgeons had with the control article was likely considerably higher than the test group. The control article was designed on an IOL platform that had been in use for more than two decades, by the largest IOL producer in the US. The test article has not been used before the approval of the test IOL (July 2022), which is relevant because of its unique optic and haptic design.
The many study strengths outweigh these weaknesses, and therefore, the study results can be considered valid
Recommendations for Labeling Changes Yes.


Lenstec SBL-3 PAS Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
6 month report 01/20/2023 01/19/2023 On Time
1 year report 07/22/2023 07/20/2023 On Time
18 month report 01/20/2024 01/19/2024 On Time
2 year report 07/21/2024 07/19/2024 On Time
final report 06/30/2025 06/27/2025 On Time