The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

Avive AED Post-Approval Study


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Completed
Application Number /
Requirement Number
P210015 / PAS001
Date Original Protocol Accepted 10/31/2022
Date Current Protocol Accepted 10/31/2022
Study Name Avive AED Post-Approval Study
Device Name Avive Automated External Defibrillator (AED) System
General Study Protocol Parameters
Study Design Prospective Cohort Study
Data Source New Data Collection
Comparison Group Objective Performance Criterion
Analysis Type Analytical
Study Population Child: 2 yrs < 12 yrs, Adolescent: 13-18 yrs, Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives This is a prospective observational study comparing the Avive AED Rhythm Recognition Detector performance to objective performance criterion.
Study Population N/A
Sample Size 200 cases
Key Study Endpoints The primary study endpoint for this PAS is a >90% sensitivity on Course VF rhythms for the Avive AED, with a 90% One-sided Lower Confidence Limit that is >87%.
The following will also be reported:
-Performance on Rapid VT rhythms for the Avive AED
-Performance on Normal Sinus rhythms for the Avive AED
-Performance on Asystole rhythms for the Avive AED
-Performance on other non-shockable rhythms for the Avive AED, including Atrial Fibrillation, Supraventricular Tachycardia, Sinus Bradycardia, Heart Block, Idioventricular rhythms, and Premature Ventricular Contractions
-Performance on Fine VF for the Avive AED
-Performance on Other VT for the Avive AED
In addition to the ECG analysis performance, the PAS will also report the usability performance of the Avive AED represented by the time elapsed from AED activation to pads placed and if shockable, time to first shock delivered. These data will be extracted and reported from the same AED device log. This information provides the most objective metric of the device usability, incorporating the critical tasks that are required to provide patient treatment.
Follow-up Visits and Length of Follow-up No follow-up required
Interim or Final Data Summary
Actual Number of Patients Enrolled 1,909 ECG rhythm segments total. 222 ECG rhythm segments with Coarse VF.
Actual Number of Sites Enrolled N/A
Patient Follow-up Rate No follow-up needed in this study. Overall, the study demonstrates a 95.5% sensitivity on Coarse VF rhythms with a 90% one-sided lower confidence limit that is 93.7% across 222 collected cases.
Final Safety Findings There were 11 instances out of 287 (3.8%) where the adjudicator consensus was either Coarse VF (10/11 instances), or Rapid VT (1/11 instances), and the AED assessed a non-shockable rhythm. None of these 11 instances had 3/3 agreement amongst the three cardiologist adjudicators. Most included a combination of shockable (Rapid VT or Coarse VF) and intermediate (Other VT or Fine VF) classifications. In 7/11 of the instances, the rhythm presented with a slow rate (<160 BPM), a low-amplitude, and semi-organized electrical activity where an Other VT, Fine VF or Pulseless Electrical Activity (PEA, which is not an option that the adjudicators can select) could have reasonably been an appropriate classification. In 10/11 of the instances, the rhythm had large deflections and/or errant QRS complexes mid-rhythm. By design, a rhythm presenting with significant indeterminate artifact can influence the AED rhythm recognition detector to not advise a shock due to safety considerations.
Final Effect Findings The results were a 95.5% sensitivity on Coarse VF rhythms for the Avive AED, with a 90% one-sided lower confidence limit that is 93.7% across 222 collected cases.
Study Strengths & Weaknesses The study collected data from real-world use of the Avive AED and demonstrated an acceptable performance for detection of Coarse VF and an acceptable time from power on to both pad placement and shock (if delivered). ECG segments seen by the Avive AED in the field are borderline rhythms, may be deteriorating from one arrythmia type to another, contain noise and artifact, and when viewed in an isolated 7-second segment can be interpreted differently even by board certified cardiologists.
Recommendations for Labeling Changes None


Avive AED Post-Approval Study Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
6 month report 05/01/2023 04/28/2023 On Time
1 year report 10/31/2023 11/01/2023 Overdue/Received
18 month report 04/30/2024 05/01/2024 Overdue/Received
2 year report 10/30/2024 10/30/2024 On Time
3 year report 10/30/2025 11/10/2025 Overdue/Received
final report 06/10/2026 06/10/2026 On Time