| General |
| Study Status |
Delayed |
Application Number / Requirement Number |
P220029 / PAS002 |
| Date Original Protocol Accepted |
08/25/2023
|
| Date Current Protocol Accepted |
12/15/2025
|
| Study Name |
PEAK PAS
|
| Device Name |
Optilume™ BPH Catheter System
|
| Clinical Trial Number(s) |
NCT04131907
|
| General Study Protocol Parameters |
| Study Design |
Prospective Cohort Study
|
| Data Source |
New Data Collection
|
| Comparison Group |
Device Subjects Serve as Own Control
|
| Analysis Type |
Analytical
|
| Study Population |
Adult: At least 22 yrs
|
| Detailed Study Protocol Parameters |
| Study Objectives |
Prospective, single-arm, multi-center, post market clinical trial to verify the continued safety and effectiveness for the Optilume BPH Catheter System
|
| Study Population |
Male subject greater than or equal to 50 years old Diagnosed with lower urinary tract symptoms (LUTS) secondary to obstructive benign prostatic hyperplasia (BPH) Prostate volume < 80 g with a prostatic urethral length between 32-55 mm as determined by trans-rectal ultrasound (TRUS)
|
| Sample Size |
Number of subjects: 92, 34 in semen sub-study Assumptions for sample size estimation: One-sided alpha of 0.025 and 80% power, based on performance goal of 30% improvement in International Prostate Symptom Score (IPSS) giving a minimum sample size of 73. Accounting for approximately 20% loss to follow-up through the full duration of the study (5 years) gives a total sample size of 92 subjects. Semen sub-study sample size is based on anticipated standard deviation of 60%, performance goal of not more than 30% decrease from baseline, and statistical power of 80%, based on one-sided 0.025 alpha level t-test for a mean for 34 subjects. Number of sites: 15 Sites location: within the United States
|
| Key Study Endpoints |
Safety Endpoints Primary: Freedom from composite treatment-related serious adverse events. Secondary (semen sub-study only): The average change from baseline in sperm concentration at 13 weeks (3 months) post-procedure will be compared against a performance goal of -30% (30% decrease from baseline). Ancillary: Frequency and severity of treatment-related adverse events. Change from baseline in semen characteristics over time (semen sub-study only). Effectiveness Endpoints Primary: Average International Prostate Symptom Score (IPSS) improvement from baseline to 12 months. Ancillary: Improvement in IPSS over time Improvement in Internation Continence Society (ICS) male short form over time Improvement in peak urinary flow (Qmax) over time Improvement in post-void residual volume (PVR) over time Freedom from repeat intervention (time-to-event)
|
| Follow-up Visits and Length of Follow-up |
Follow-up visits at 1-, 3-, 6-, and 12-months post-procedure, and annually thereafter through 5 years
|
| Interim or Final Data Summary |
| Actual Number of Patients Enrolled |
165
|
| Actual Number of Sites Enrolled |
7
|